pH Electrodes Instruction Sheet
Models 601100, 601500 and 60120B
Electrode Specifications
Model number 601100 60120B 601500
Description Flat surface combination pH
electrode with gelled reference
solution
Mini combination pH electrode with
solution and detachable bulb guard
General-purpose combination pH
electrode with sealed gelled reference
filling solution and detachable bulb guard
Dimensions 15 x 115 mm 10 x 126 mm 12.5 x 155 mm
Construction material Polymer housing, flat glass bulb Polymer housing, glass bulb Polymer housing, glass bulb
Measuring range 0 to 14.0 pH units 0 to 14.0 pH units 0 to 14.0 pH units
Temp. range 0-80
o
C 0-80
o
C 0-80
o
C
Reference junction Teflon Ceramic Ceramic
Reference type Ag/AgCl (silver silver chloride) Ag/AgCl (silver silver chloride)Ag/AgCl (silver silver chloride)
Isopotential point ~7.0 pH ~7.0 pH ~7.0 pH
Precision 0.02 pH units 0.02 pH units 0.02 pH units
Connector BNC BNC BNC
Cable length 1 meter (39”) 1 meter (39) 1 meter (39”)
Storage container Bottle Cap Bottle
General Electrode Procedure
1. To avoid damage to the electrode when removing it from or
inserting it into the soaking bottle;
a) Unscrew the bottle cap.
b) Remove the electrode and cap from the bottle.
c) Slide the cap off the electrode.
d) Reverse this procedure when storing the electrode.
2. Rinse the electrode with deionized or distilled water.
3. Carefully shake electrode to ensure that any air bubbles in the
bulb or lower stem of the electrode are dislodged upwards.
4. Blot electrode with tissue.
5. Connect electrode to pH meter and follow the pH meter
manufacturer’s instruction for calibration.
Before Measurement
1. Generally, the sample should be aqueous and fall within the
range 0 – 14 pH
2. Always use fresh buffers for calibration. Choose buffers that are
no more than 3 pH units apart.
3. To obtain the maximum precision, always buffer as close as
possible to the expected measured value.
4. Between measurements, rinse electrodes with distilled water and
then with the next solution to be measured.
5. Stir all buffers and samples.
6. Avoid rubbing or wiping electrode bulb in order to reduce the
chance of error due to polarization.
pH Calibration and Measurement
Refer to the owner’s guide from the manufacturer of the pH meter
being used for detailed information regarding calibration and
temperature compensation procedures.
Two-Buffer Calibration (for high precision measurements)
1. Set up the pH meter according to the manufacturer’s instruction
manual.
2. Ensure that all buffers are at the same temperature (+/- 2
o
C.) If
samples are at varying temperatures, temperature compensation
is recommended. (See pH meter instruction manual.)
3. Select two buffers that bracket the expected pH value. The first
should be near the electrode isopotential point (pH 7) and the
second near the expected pH of the sample. (e.g., pH 4 or 10.)
4. Rinse the electrode first with distilled water and then with pH 7
buffer. Place the electrode in the pH 7 buffer.
5. Wait for a stable pH 7 reading. Set the pH meter to the pH value
of the buffer at its measured temperature.
6. Rinse the electrode first with distilled water and then with the
second buffer. Place the electrode in the second buffer.
7. Wait for a stable display. Set the pH meter to the pH value of the
buffer at its measured temperature.
8. Calibration is complete. Proceed to pH measurement
Single-Buffer Calibration (for lower precision measurements)
1. Set up the pH meter according to the manufacturer’s instruction
manual.
2. Choose a buffer that has a value close to the expected pH of the
sample.
3.
Rinse the electrode first in distilled water and then in the buffer being used
for calibration. Place the electrode in the buffer.
4. Wait for a stable display. Set the pH meter to the pH value of the buffer at
its measured temperature.
5. Proceed to pH measurement
pH Measurement
1. Calibrate the electrode as described in previous section.
2. Rinse the electrode with distilled water and then with the sample solution.
3. Place the electrode in the sample.
4. Stir the sample.
5. When the reading is stable, record the pH value.
Electrode Storage and Maintenance
Electrode Storage
1. Electrodes 601100 and 601500
Store in the soaking bottle. The soaking solution should be 22.4g
KCl dissolved in 100ml pH4 buffer solution or 25g KCl dissolved
in 100ml pure water.
2. Electrode 60120B
Store with wetting cap kept damp with pH4 buffer or, alternatively,
store in a soaking solution.
Electrode Maintenance
3. Inspect the electrode for scratches, cracks, salt crystal build-up,
or membrane/junction deposits.
4. Rinse off any salt build-up with distilled water, and remove any
membrane/junction deposits as directed in cleaning procedures
below.
5. Drain the reference chamber, flush it with fresh filling solution and
refill the chamber
Cleaning Procedures
Soak in mild detergent or dilute 0.1M HCl for 30 minutes
Electrode Reconditioning
If the electrode has been allowed to dry for any period of time it
may require reconditioning. Soak the electrode in a pH 4 buffer
solution for at least 20 minutes. Depending on the dryness of the
reference junction, this reconditioning process may require up to a
24 hour soaking period in order to obtain accurate pH readings.
Warranty
FLIRSystems, InstrumentsInc.warrantsthisExtech branddevice to be free ofdefects
inpartsand workmanship for oneyear from date of shipment (a six month limited
warrantyappliestosensorsandcables).Ifitshouldbecomenecessarytoreturnthe
instrumentforserviceduringorbeyond the warranty
period, contact the Customer
ServiceDepartmentforauthorization.Visitthewebsitewww.extech.com for contact
information.A Return Authorization (RA) number must be issued beforeanyproduct is
returned.Thesenderisresponsibleforshippingcharges,freight, insurance and proper
packagingtopreventdamageintransit.Thiswarrantydoesnotapplytodefects
resultingfromactionoftheusersuchasmisuse,improperwiring,operationoutside
of
specification,impropermaintenanceor orrepair, unauthorizedmodification.FLIR
Systems,Inc.specificallydisclaims any implied warranties or merchantability or fitness
for a specificpurposeandwillnot be liable for any direct, indirect, incidentalor
consequentialdamages.FLIR’s total liability is limited to repair or replacementofthe
product.
Thewarrantysetforthaboveis inclusive andnootherwarranty,whether
written or oral, is expressed or implied.


Produkt Specifikationer

Mærke: Extech
Kategori: Måling
Model: 601500

Har du brug for hjælp?

Hvis du har brug for hjælp til Extech 601500 stil et spørgsmål nedenfor, og andre brugere vil svare dig




Måling Extech Manualer

Extech

Extech HDV-TX2L Manual

2 December 2025
Extech

Extech MG325 Manual

1 December 2025
Extech

Extech TL900 Manual

1 December 2025
Extech

Extech 601500 Manual

1 December 2025
Extech

Extech IRT600 Manual

1 December 2025
Extech

Extech RH35 Manual

1 December 2025
Extech

Extech CA3010 Manual

1 December 2025
Extech

Extech MO297 Manual

1 December 2025
Extech

Extech MA435T Manual

1 December 2025
Extech

Extech CL205 Manual

1 December 2025

Måling Manualer

Nyeste Måling Manualer

Vevor

Vevor EM4556 Manual

1 December 2025
Vevor

Vevor SD317 Manual

1 December 2025
Vevor

Vevor SD308 Manual

1 December 2025
Seaward

Seaward PV:1525 Manual

30 November 2025
FLIR

FLIR E75 Manual

28 November 2025
Terraillon

Terraillon COACH Manual

27 November 2025
FLIR

FLIR T530 Manual

27 November 2025
FLIR

FLIR TG268 Manual

27 November 2025