Model 705
Loop Calibrator
Calibration Information
W Warning
To avoid electrical shock, remove test leads and any input signals from the
Model 705 Loop Calibrator before opening the case.
Caution
The Model 705 Loop Calibrator contains parts that can be damaged by static
discharge. Follow the standard practices for handling static sensitive
devices.
Introduction
The Calibration Information for the Model 705 Loop
Calibrator (hereafter referred to as the Loop
Calibrator) provides the information necessary to
calibrate and verify the performance of the Loop
Calibrator.
This document includes the following:
Contacting Fluke for service
Specifications
Cleaning
Replacing the battery
Disassembly and reassembly
Recommended test equipment
Performance verification procedures
Calibration adjustments
User-replaceable parts
Complete operating instructions and the warranty
statement for the Model 705 Loop Calibrator are
provided in the that came with it.Instruction Sheet
Contacting Fluke for Service
This Fluke product will be free from defects in
material and workmanship for three years from the
date of purchase.
The complete warranty statement is in the Model 705
Loop Calibrator Instruction Sheet.
To locate an authorized service center, visit us on the
World Wide Web: or call Fluke usingwww.fluke.com
the phone numbers listed below:
1-888-99FLUKE (1-800-993-5853) in USA
1-800-36-FLUKE (1-800-363-5853) in Canada
+31 402-678-200 in Europe
+81-3-3434-0181 in Japan
+65-738-5655 in Singapore
+1-425-356-5500 from other countries
Fluke 705
Calibration Information
2
Specifications
Product specifications are provided Tables 1 and 2. These specifications are based on a 1-year calibration cycle and
apply from 18 ºC to 28 ºC (64 ºF to 82 ºF) unless stated otherwise. Accuracy specifications are given as follows:
± ( [ % of reading ] + [ number of least significant digits ] ).
Table 1. General Specifications
Maximum voltage between any
terminal and earth ground or
between any two terminals
30 V
Display: 0 to 28 V with 0.001 V resolution, 0 to 24 mA with 0.001 mA resolution
Operating Temperature: -10 °C to 55 °C
Storage Temperature: -40 ° °C to 60 C
Temperature Coefficient: ± ° ° 0.005 % of range per C for -0 C to 18 C to 55 °C and 28 ° °C
Relative Humidity: 95 % to 30 °C,
75 % to 40 °C,
45 % up to 50 °C,
35 % up to 55 °C
Operationing Altitude: 3000 meters maximum
Power Requirements: Single 9 V battery (ANSI/NEDA 1604A or IEC 6LR61)
Battery Life (Typical): Source Mode: 18 hours; 12 mA into 500
Measure / Simulate Mode: 50 hours
Shock, Vibration: 1 meter drop test
Random vibration 2 g @ 5 to 500 Hz
Size (H x W x L): 32 mm x 87 mm x 187 mm (1.25 in x 3.41 in x 7.35 in)
Weight: 349 g (12.3 oz). With Holster and flex-stand: 601 g (21.2 oz)
Safety Certified as compliant to CAN/CSA C22.2 No. 1010.1:1992.
Complies with ANSI/ISA S82.01-1994
Certification P and )
Table 2. Accurcy Specifications
Function Range Resolution Accuracy
DC Volts Input + 28 V 0.001 V ± (0.025 % + 1)
DC mA Input 24 mA 0.001 mA ±( 0.02 % + 2)
DC mA Output 0 mA to 24 mA 0.001 mA ± (0.02 % + 2)
E 0 % -100 % - 0 % 40 second smooth ramp
O 0 % -100 % - 0 % 15 second smooth ramp
N 0 % -100 % - 0 % Stair-step ramp in 25 % steps, pausing 5 seconds at each step
Source mode: Compliance: ≥ Ω 1000 at 20 mA
Simulate mode: External loop voltage requirements: 24 V nominal, 30 V maximum, 12 V minimum
Loop Power 24 V
Percent display - 25 % to 125 %
Input/Output protection Input/Output protection on mA range: resettable, nonreplaceable 0.1 A fuse.


Produkt Specifikationer

Mærke: Fluke
Kategori: Ikke kategoriseret
Model: 705
Bredde: 70 mm
Dybde: 41 mm
Højde: 152 mm
Vægt: 340 g
Brugervejledning: Ja
Kontroltype: Knapper
Batterispænding: 9 V
Opbevaringstemperatur (T-T): -40 - 60 °C
Relativ luftfugtighed ved drift (H-H): 0 - 90 %
Driftstemperatur (T-T): -10 - 55 °C
Spændingsmåling opløsning: 0.001 V
Hylster inkluderet: Ja
Krokodille clips: Ja
Kontrol af spænd: Ja
Automatisk trin: Ja
Ringspænding: 30 V
Spændingsmåling nøjagtighed: 0.25 %

Har du brug for hjælp?

Hvis du har brug for hjælp til Fluke 705 stil et spørgsmål nedenfor, og andre brugere vil svare dig




Ikke kategoriseret Fluke Manualer

Fluke

Fluke 1555 Manual

18 Juni 2025
Fluke

Fluke SB140 Manual

14 Juni 2025
Fluke

Fluke Pro3000 Manual

11 Marts 2025
Fluke

Fluke 438-II Manual

8 Oktober 2024
Fluke

Fluke 1670 Manual

8 Oktober 2024
Fluke

Fluke ii905 Manual

8 Oktober 2024
Fluke

Fluke ii915 Manual

8 Oktober 2024
Fluke

Fluke ii500 Manual

7 Oktober 2024
Fluke

Fluke T5-1000 Manual

4 Oktober 2024

Ikke kategoriseret Manualer

Nyeste Ikke kategoriseret Manualer