Model 707
Loop Calibrator
Calibration Information
W�Warning
To prevent electrical shock, remove test leads and any input signals from the
Model 707 Loop Calibrator before opening the case.
WCaution
The Model 707 Loop Calibrator contains parts that can be damaged by static
discharge. Follow the standard practices for handling static sensitive
devices.
Introduction
The for the Model 707 Loop Calibration Information
Calibrator (hereafter referred to as the Loop
Calibrator) provides the information necessary to
verify the performance and adjust the calibration of
the Loop Calibrator.
This document includes the following:
Contacting Fluke for service
Specifications
Cleaning
Replacing the battery
Disassembly and reassembly
Recommended test equipment
Performance verification procedures
Calibration adjustments
User-replaceable parts
The complete operating instructions and warranty
statement for the Model 707 Loop Calibrator are
provided in the Instruction Sheet that came with it.
Contacting Fluke for Service
This Fluke product will be free from defects in
material and workmanship for three years from the
date of purchase.
To contact Fluke, call one of the following telephone
numbers:
Technical Support USA: 1-800 44 800 443 5853)- - -FLUKE (1 - -
Calibration/Repair USA: 1-888 99 888 993 5853)- - -FLUKE (1 - -
Canada: 1-800 36 800 363 5853)- - -FLUKE (1 - -
Europe: +31 402-675 200-
Japan: +81-3-6714 3114-
Singapore: +65-6799 5566-
China: +86-400 921 0835- -
Brazil: +55- - - 11 3530 8901
Anywhere in the world: +1-425 446 5500- -
Or, visit Fluke's website at www.fluke.com.
To register your product, visit http://register.fluke.com.
To view, print, or download the latest manual supplement,
visit http://us.fluke.com/usen/support/manuals.
Address correspondence to:
Fluke Corporation Fluke Europe B.V.
P.O. Box 9090, P.O. Box 1186,
Everett, WA 98206-9090 5602 BD Eindhoven
U.S.A. The Netherland
Fluke 707
Calibration Information
2
Specifications
Product specifications are provided in Tables 1 and 2. These specifications are based on a 1-year calibration cycle
and apply from 18 ºC to 28 ºC (64 ºF to 82 ºF) unless stated otherwise. Accuracy specifications are given as
follows:
± ( [ % of reading ] + [ number of least significant digits ] ).
Table 1. General Specifications
Maximum voltage between any
terminal and earth ground or
between any two terminals
30 V
Display: 0 to 28 V with 0.001 V resolution, 0 to 24 mA with 0.001 mA resolution
Operating Temperature: -10 °C to 55 °C
Storage Temperature: -30 °C to 60 °C
Temperature Coefficient: ± 0.005 % of range per C to 18 d 28 C to 55 °C for -10 ° °C an ° °C
Relative Humidity: 95 % to 30 °C,
75 % to 40 °C,
45 % up to 50 °C,
35 % up to 55 °C
Operating Altitude: 3000 meters maximum
Power Requirements: Single 9 V battery (IEC 6LR61)
Battery Life (Typical): Source Mode: 18 hours; 12 mA into 500
Measure / Simulate Mode: 50 hours
Size (H x W x L):
With Holster:
50.80 mm x 69.85 mm x 142.87 mm (2.00 in x 2.75 in x 5.625 in)
54.61 mm x 76.20 mm x 158.75 (2.15 in x 3.00 in x 6.25 in)
Weight: 349 g (12.3 oz). With Holster and flex- stand: 601 g (21.2 oz)
Safety:
IEC 61010 1: Pollution degree 2-
IEC 61010 030: CAT O 30-2- V
Table 2. Accuracy Specifications
Function Range Resolution
Accuracy
(% of reading + counts)
DC Volts Input
+ 28 V (+30 V max) 0.001 V ± (0.015 % + 2 counts)
DC mA Input 20 mA (24 mA max) 0.001 mA ± (0.015 % + 2 counts)
DC mA Output 0 mA to 24 mA 0.001 mA ± (0.015 % + 2 counts)
Source mode: Compliance: ≥ Ω 1200 at 20 mA, 950 at 20 mA in Hart Mode
Simulate mode: External loop voltage requirements: 24 V nominal, 28 V maxi 2 V minimummum, 1
Loop Power 24 V
Percent display
- 25 % to 125 % (4 to 20 mA mode)
Input/Output protection Fuseless protection.


Produkt Specifikationer

Mærke: Fluke
Kategori: Ikke kategoriseret
Model: 707

Har du brug for hjælp?

Hvis du har brug for hjælp til Fluke 707 stil et spørgsmål nedenfor, og andre brugere vil svare dig




Ikke kategoriseret Fluke Manualer

Fluke

Fluke 1555 Manual

18 Juni 2025
Fluke

Fluke SB140 Manual

14 Juni 2025
Fluke

Fluke Pro3000 Manual

11 Marts 2025
Fluke

Fluke 438-II Manual

8 Oktober 2024
Fluke

Fluke 1670 Manual

8 Oktober 2024
Fluke

Fluke ii905 Manual

8 Oktober 2024
Fluke

Fluke ii915 Manual

8 Oktober 2024
Fluke

Fluke ii500 Manual

7 Oktober 2024
Fluke

Fluke T5-1000 Manual

4 Oktober 2024

Ikke kategoriseret Manualer

Nyeste Ikke kategoriseret Manualer