Fluke 80PK-27 Manual


Læs gratis den danske manual til Fluke 80PK-27 (5 sider) i kategorien Ikke kategoriseret. Denne vejledning er vurderet som hjælpsom af 23 personer og har en gennemsnitlig bedømmelse på 3.7 stjerner ud af 12 anmeldelser. Har du et spørgsmål om Fluke 80PK-27, eller vil du spørge andre brugere om produktet?

Side 1/5
®
80PK-27
SureGripTM Industrial Surface
Temperature Probe
Instruction Sheet
XWWarning
To avoid electrical shock, do not use this probe
when voltages exceeding 24 V ac rms or 60 V dc
are present. The probe tip is electronically
connected to the output terminals.
XW Attention
Pour éviter les risques d’électrocution, ne pas
utiliser cette sonde quand la tension est
sepérieure à 24 V efficaces en courant alternatif
ou 60 V en courant continu. La pointe de la sonde
est reliée électriquement aux bornes de sorties.
XW Warnung
Um Elektroschock zu vermeiden, dürfen Sie den
Meßstift nicht in der Anwesenheit von
Spannungen über 24V Wechselstrom oder 60V
Gleichstrom verwenden. Die Spitze des Meßstifts
ist mit den Ausgangsanschlüssen elektrisch
verbunden.
XW Advertencia
Para evitar descargas eléctricas, no use esta
sonda cuando se apliquen voltajes que
sobrepasen 24 V ca rcm o 60 V cc. La punta de la
sonda está conectada eléctricamente a las
terminales de salida.
Introduction
The 80PK-27 SureGrip
TM Industrial Surface Temperature
Probe is designed for reliably measuring temperatures to 600
° °C (1112 F) on flat or slightly convex surfaces, while retaining
a long service life.
Specifications
Type: K (Chromel vs Alumel)
Measurement Range: -127 C °C to 600 °
(-196.6 °F to 1112 °F)
Junction Accuracy: -127 C °C to 0 °C ±4.4 °
(-196.6 °F to 32 °F ± 7.92 °F)
0 °C to 183 °C ±2.2 °C
(32 °F to 361 °F ±4.86 °F)
183 °C to 600 °C ±1.2 % of reading in °C
(361 °F to
1112 °F ± <1.2 % of reading
in °F)
Output: 25 ° ° °C (77 F) = 1.00 mV (Reference junction at 0 C)
Seebeck Coefficient: 25 °C (77 °F) = 40.50 µ °V/ C
November 2004
© 2004 Fluke Corporation. All Rights Reserved.
All product names are trademarks of their respective companies
Measurement Time (Time Constant): 1 time constant equals
330 ms; 5 time constants equal final reading. See “Application
Information” below.
Maximum Voltage Rating at Probe Tip: 24 V ac rms, or 60 V
dc
Probe Tip: Maximum Temperature: 600 °C
(1112 °F)
Material: 303 stainless steel
Sheath Material: 303 stainless steel
Cable Length: 40 inches (1 meter)
Cable Isulation: Material: PVC
Maximum Temperature: 105 °C (220 °F)
Minimum Temperature: -20 °F
Conductors: Type: K
Size: AWG #24 stranded (7 strands of #32)
Length: 1.2 meters nominal (4 ft)
Handle: Material: Hytrel
Maximum Temperature: 65.6 °C (150 °F)
Minimum Temperature: -40 °C (-40 °F)
Connector:
Type: yellow mini-thermocouple connector with 0.792
mm (0.312 in) pin spacing
Material: Hytrel
Maximum Temperature: 125 °C (275 °F)
Dimensions:
Diameter: 22.9 mm (0.9 in)
Length: 322.6 mm (12.75 in)
Measurement Considerations
Instrument Compatibility
The 80PK-27 is designed to be compatible with any
temperature measuring instrument that accepts Type K
thermocouples, has a miniature thermocouple connector, and
has cold reference junction compensation. Accuracy of the
temperature measuring instrument must be considered along
with the 80PK-27 accuracy specifications in order to determine
the overall accuracy of the combination.
Temperature Limitations
The probe tip has continuous temperature rating of 600 °C.
However the rest of the assembly is rated for a lower
temperature. See the specifications for further information.
Media Limitations
Media must be compatible with Chromel, Alumel, and 303
stainless steel.
Application Information
At high temperatures, a surface temperature probe removes a
small amount of heat from the measured surface. At 600 °C on
a polished metal surface, the temperature at a contact point
will be lowered, typically not more than 2 °C. A lowering of the
temperature at the contact point is less likely (and contact
response time is quicker) on clean, polished, thermally
conductive surfaces than on materials with low thermal
conductivity, such as plastic or rough, contaminated surfaces.
To obtain the best thermal contact and performance, the
stainless steel ring must make full and firm contact with the
measurement surface.


Produkt Specifikationer

Mærke: Fluke
Kategori: Ikke kategoriseret
Model: 80PK-27
Antal pr. pakke: 1 stk
Produktfarve: Black, Silver, Yellow
Produkttype: Test sonde
Ledningslængde: 1 m
Materiale: Rustfrit stål
Isoleringsmateriale: PVC
Længde: 322.6 mm
Jævnspænding (maks.): 60 V

Har du brug for hjælp?

Hvis du har brug for hjælp til Fluke 80PK-27 stil et spørgsmål nedenfor, og andre brugere vil svare dig




Ikke kategoriseret Fluke Manualer

Fluke

Fluke 1555 Manual

18 Juni 2025
Fluke

Fluke SB140 Manual

14 Juni 2025
Fluke

Fluke Pro3000 Manual

11 Marts 2025
Fluke

Fluke 438-II Manual

8 Oktober 2024
Fluke

Fluke 1670 Manual

8 Oktober 2024
Fluke

Fluke ii905 Manual

8 Oktober 2024
Fluke

Fluke ii915 Manual

8 Oktober 2024
Fluke

Fluke ii500 Manual

7 Oktober 2024
Fluke

Fluke T5-1000 Manual

4 Oktober 2024

Ikke kategoriseret Manualer

Nyeste Ikke kategoriseret Manualer