Læs gratis den danske manual til Fluke i800 AC (6 sider) i kategorien Ikke kategoriseret. Denne vejledning er vurderet som hjælpsom af 33 personer og har en gennemsnitlig bedømmelse på 4.8 stjerner ud af 17 anmeldelser. Har du et spørgsmål om Fluke i800 AC, eller vil du spørge andre brugere om produktet?

Side 1/6
PN 2842289
June 2007 Rev.1, 8/08
©2007-2008 Fluke Corporation. All rights reserved. Product
specifications are subject to change without notice. Printed in
France.
®
i800
AC Current Clamp
Instruction Sheet
Introduction
The i800 AC Current Clamp, the Clamp, has been designed for use
with multimeters, recorders, power analyzers, safety testers, etc., for
accurate non-intrusive ac current measurement.
Using the latest transformer technology, the Clamp measures
currents from 100 mA to 800 A.
The Clamp complies with the safety standards EN61010-2-032 for
input and output circuits.
Symbols
The table below lists the symbols used on the Clamp and/or in this
manual.
Symbol Description
~ Do not dispose of this product as unsorted municipal
waste. Go to Fluke’s website for recycling information.
W Important Information. See manual.
X Hazardous Voltage. Risk of electric shock.
T Double insulation.
, Application around and removal from HAZARDOUS LIVE
conductors is permissible.
) Conforms to Canadian Standards Association.
P Complies with the relevant European standards.
; Conforms to Australian standards.
CAT III
Equipment designed to protect against transients in
equipment in fixed-equipment installations, such as
distribution panels, feeders and short branch circuits, and
lighting systems in large buildings.
Safety Instructions
Please read this section carefully. It will make you familiar with the
most important safety instructions for handling the Clamp. In this
instruction sheet, a
Warning identifies conditions and actions that
pose hazard(s) to the user. A
Caution identifies conditions and
actions that may damage the Clamp or the test instruments.
WX Warning
The Clamp may only be used and handled by qualified
personnel. To avoid electric shock or personal injury,
follow these precautions:
Do not use the Clamp if damaged. Always
connect to display device before it is installed
around the conductor.
If the Clamp is not used as specified by the
manual, protection provided by the Clamp may
be impaired.
Do not clamp around a conductor before
connecting clamp output to the measuring
equipment.
Do not disconnect the output from the measuring
equipment while the Clamp is still around a
conductor.
Connect the Clamp output to measuring
equipment having safety receptacles.
Do not use prior to consulting local and national
safety codes and use of protective equipment is
recommended.
Do not hold the Clamp beyond the tactile barrier.
Before each use, inspect the Clamp. Look for
cracks or missing portions of the housing or
output cable insulation. Also look for loose or
weakened components. Pay particular attention
to the insulation surrounding the jaws. Open and
close the Clamp, noting detent action indicating
proper mechanical operation of SIAC
TM anti short
circuit feature.
Use caution when working with voltages above
60 V dc, 30 V ac rms or 42 V ac peak. Such
voltages pose a shock hazard.
Do not use Clamp in wet environments or in
locations that hazardous gases exist.


Produkt Specifikationer

Mærke: Fluke
Kategori: Ikke kategoriseret
Model: i800 AC

Har du brug for hjælp?

Hvis du har brug for hjælp til Fluke i800 AC stil et spørgsmål nedenfor, og andre brugere vil svare dig




Ikke kategoriseret Fluke Manualer

Fluke

Fluke 1555 Manual

18 Juni 2025
Fluke

Fluke SB140 Manual

14 Juni 2025
Fluke

Fluke Pro3000 Manual

11 Marts 2025
Fluke

Fluke 438-II Manual

8 Oktober 2024
Fluke

Fluke 1670 Manual

8 Oktober 2024
Fluke

Fluke ii905 Manual

8 Oktober 2024
Fluke

Fluke ii915 Manual

8 Oktober 2024
Fluke

Fluke ii500 Manual

7 Oktober 2024
Fluke

Fluke T5-1000 Manual

4 Oktober 2024

Ikke kategoriseret Manualer

Nyeste Ikke kategoriseret Manualer