Fluke T5-1000 Manual


Læs gratis den danske manual til Fluke T5-1000 (9 sider) i kategorien Ikke kategoriseret. Denne vejledning er vurderet som hjælpsom af 62 personer og har en gennemsnitlig bedømmelse på 4.4 stjerner ud af 31.5 anmeldelser. Har du et spørgsmål om Fluke T5-1000, eller vil du spørge andre brugere om produktet?

Side 1/9
PN 686953
March 1998 Rev.1, 9/01
1998-2001 Fluke Corporation. All rights reserved. Printed in U.S.A. Product names are trademarks of their respective companies.
T5-600/T5-1000
Electrical Tester
Service Information
Introduction
This service information sheet provides the following information for the T5-600 and T5-1000 Electrical
Testers (hereafter referred to as "the tester").
Safety information
Parts and service information
Specifications
Cleaning procedure
Required equipment
Performance tests
Parts and accessories list
Battery replacement procedure
For operating instructions, refer to the T5-600/T5-1000 Electrical Tester Instruction Sheet.
Safety Information
Warning
To avoid possible electric shock or personal injury, follow these guidelines:
Do not use the tester if it is damaged or operating abnormally. Protection
may be impaired.
Before each use:
Make sure the battery door is closed and latched.
Inspect the tester and test leads. Look for cracks, missing plastic, exposed
metal, or damaged insulation. Replace damaged test leads before using the
tester.
Verify the tester’s operation by measuring a known voltage.
Replace the batteries as soon as the low battery indicator (M) appears.
Do not use the tester around explosive gas, vapor or dust.
Do not apply more than the rated voltage, as marked on the tester, between
terminals or between any terminal and earth ground.
Refer servicing to qualified personnel.
Use caution when working above 30 V ac rms, 42 V ac peak, or 60 V dc.
When using the probes, keep your fingers behind the finger guards on the
probes.
Connect the common test lead before you connect the live test lead.
Disconnect the live test lead first.
The display will not display hazardous voltages in display hold. The voltage
indicator LED ( ) continues to operate.
®
T5-600/T5-1000
Electrical Tester
2
Parts and Service
The tester is warranted to be free from defects in material and workmanship for two years, while under normal
use. Parts and repairs are warranted for 90 days. For the complete warranty statement, refer to the
T5-600/T5-1000 Electrical Tester Instruction Sheet.
To order parts, or for warranty service, contact Fluke as follows:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke’s Web site at www.fluke.com.
Specifications
Accuracy is specified for one year after calibration, at 18 °C to 28 °C (64 °F to 82 °F) with relative humidity
to 90 %. AC conversions are ac-coupled, average responding, and calibrated to the rms value of a sine wave
input. Accuracy specifications are given as follows:
±([ % of reading] + [number of least significant digits])
Temperature coefficient of 0.1 x (specified accuracy)/ °C for <18 °C or >28 C (<64.4 ° °F or >82.4 °F)
Serial Numbers 79560000 and Above
General Specifications
Calibration One-year calibration cycle.
Maximum Voltage
Between any Terminal
and Earth Ground
T5-600: 600 V rms, Overvoltage Category III, Pollution Degree 2
T5-1000: 1000 V rms, Overvoltage Category III, 600 V RMS, OverVoltage Category IV,
Pollution Degree 2
Maximum Voltage
Between Current Fork
and Earth Ground
T5-600 and T5-1000: 1000 V rms, Overvoltage Catergory III, also 600 V rms, Overvoltage
Category IV, Pollution Degree 2, (This dual rating applies to the current fork only.)
Temperature Operating: -10 ° ° ° ° ° ° ° °C to +50 C (14 F to 122 F); Storage: -30 C to +60 C (-22 F to +140 F)
Altitude Operating: 2000 m (6562 ft); Storage: 10,000 m (32808 ft)
Relative Humidity 0 % to 95 %, 5 ° °C to 30 C (41 C (86 °F to 86°F); 0 % to 75 %, 30 ° °C to 40 °F to 104 °F);
0 % to 45 %, 40 ° °C to 50 C (104 °F to 122 °F)
Battery Type and Life AA (2); 360 hours continuous with alkaline; 125 hours continuous with zinc chloride
Shock, Vibration 1 m drop at -10 °C to + 50 °C (14 °F to 122 °F) per ANSI/ISA-S82.01-1994 and EN 61010-1
1995. Random vibration per MIL-PRF-28800F for a Class 2 instrument
(5 Hz to 55 Hz, 3 g maximum)
Surge Protection T5-600: 6 kV per IEC 61010, T5-1000: 8 kV per IEC 61010
Enclosure Rating IP 52 per IEC 60529, no vacuum applied


Produkt Specifikationer

Mærke: Fluke
Kategori: Ikke kategoriseret
Model: T5-1000

Har du brug for hjælp?

Hvis du har brug for hjælp til Fluke T5-1000 stil et spørgsmål nedenfor, og andre brugere vil svare dig




Ikke kategoriseret Fluke Manualer

Fluke

Fluke 1555 Manual

18 Juni 2025
Fluke

Fluke SB140 Manual

14 Juni 2025
Fluke

Fluke Pro3000 Manual

11 Marts 2025
Fluke

Fluke 438-II Manual

8 Oktober 2024
Fluke

Fluke 1670 Manual

8 Oktober 2024
Fluke

Fluke ii905 Manual

8 Oktober 2024
Fluke

Fluke ii915 Manual

8 Oktober 2024
Fluke

Fluke ii500 Manual

7 Oktober 2024
Fluke

Fluke T5-1000 Manual

4 Oktober 2024

Ikke kategoriseret Manualer

Nyeste Ikke kategoriseret Manualer