Mean Well RPS-300-24-C Manual


Læs gratis den danske manual til Mean Well RPS-300-24-C (4 sider) i kategorien Ikke kategoriseret. Denne vejledning er vurderet som hjælpsom af 33 personer og har en gennemsnitlig bedømmelse på 5.0 stjerner ud af 17 anmeldelser. Har du et spørgsmål om Mean Well RPS-300-24-C, eller vil du spørge andre brugere om produktet?

Side 1/4
Installation Manual
Type
Medical Switching Power Supply
(1) Enclosed (Families: MSP)
(2) Modular (Families: MP, NMP)
(3) PCB (Families: MPS, MPD, MPT, MQ, RPS, RPD, RPT, RPTG, NEM, PM)
(4) Adaptor (Families: GSM, MES)
Installation
(1) Enclosed Type
(a) Before any installation or maintenance work, please disconnect your system from the utility.
Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient
temperature may increase the internal component temperature and will require a de-rating in
output current. Please refer to the specification sheets to receive the optimum mounting
position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance
must be kept when the adjacent device is a heat source.
(d) Keep enough insulation distance between mounting screws and internal components of
Enclosed units. Please refer to case drawing on specifications to receive the maximum length of
mounting screw.
(e) Recommended wires of Enclosed units are shown as below.
AWG 18 16 14 12 10 8
Rated Current of
Equipment (Amp) 6A 6-10A 10-16A 16-25A 25-32A 32-40A
Cross-section of
Lead(mm2 ) 0.75 1.00 1.5 2.5 4 6
Note: The suggested maximum current above only suit for 1-4 strips, 5 or more
strips should not exceed 80% of current rating.
Make sure that all strands of each stranded wire enter the terminal connection and the screw
terminals are securely fixed to prevent poor contact.
(f) For other information about the products, please refer to www.meanwell.com for details.
(2) Modular Type
(a) Before any installation or maintenance work, please disconnect your system from the utility.
Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient
temperature may increase the internal component temperature and will require a de-rating in
output current. Please refer to the specification sheets to receive the optimum mounting
position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance
must be kept when the adjacent device is a heat source.
(d) Keep enough insulation distance between mounting screws and internal components of
Modular units. Please refer to case drawing on specifications to receive the maximum length of
mounting screw.
(e) Recommended wires of Modular units are shown as below.
AWG 18 16 14 12 10 8
Rated Current of
Equipment (Amp) 6A 6-10A 10-16A 16-25A 25-32A 32-40A
Cross-section of
Lead(mm2 ) 0.75 1.00 1.5 2.5 4 6
Note: The suggested maximum current above only suit for 1-4 strips, 5 or more
strips should not exceed 80% of current rating.
Make sure that all strands of each stranded wire enter the terminal connection and the screw
terminals are securely fixed to prevent poor contact.
(f) For other information about the products, please refer to www.meanwell.com for details.


Produkt Specifikationer

Mærke: Mean Well
Kategori: Ikke kategoriseret
Model: RPS-300-24-C

Har du brug for hjælp?

Hvis du har brug for hjælp til Mean Well RPS-300-24-C stil et spørgsmål nedenfor, og andre brugere vil svare dig




Ikke kategoriseret Mean Well Manualer

Ikke kategoriseret Manualer

Nyeste Ikke kategoriseret Manualer

Waterstone

Waterstone 7800 Manual

11 December 2025
Kwikset

Kwikset 258 Manual

11 December 2025
Schlage

Schlage F170 Manual

11 December 2025
Moen

Moen Align 7565EV Manual

11 December 2025
ALNO

ALNO MC10244 Manual

11 December 2025
JET

JET TAK-ZX Manual

11 December 2025
Waterstone

Waterstone 1455 Manual

11 December 2025
Laica

Laica NE2015 Manual

11 December 2025