Mean Well RPT-75D Manual


Læs gratis den danske manual til Mean Well RPT-75D (5 sider) i kategorien Ikke kategoriseret. Denne vejledning er vurderet som hjælpsom af 13 personer og har en gennemsnitlig bedømmelse på 4.2 stjerner ud af 7 anmeldelser. Har du et spørgsmål om Mean Well RPT-75D, eller vil du spørge andre brugere om produktet?

Side 1/5
72 Reliable Triple Output Medical Grade W
RPT- 75 series
Model: RPT- D Test Report 1 5 75 /
MODEL RPT- D
75
OUTPUT FUNCTION TEST
NO
TEST ITEM
SPECICATION
TEST CONDITION
VERDICT
1
RIPPLE & NOISE
V1: 80
V2: 0 20
V3: 120
mVp-p (Max )
mVp-p (Max )
mVp-p (Max )
I/P: 230VAC
O/P:FULL LOAD
Ta:25
V1:
V2:
V3:
15
26
16
mVp-p (Max )
mVp-p (Max )
mVp- (Max ) p
P
2
OUTPUT VOLTAGE
ADJUST RANGE
CH1: 4.75 V~-5.5 V
I/P: 230 VAC
I/P: 115 VAC
O/P:MIN LOAD
Ta:25
4.55 V~
4.55 V~
5.88
5.88
V / 230VAC
V / 115VAC
P
3
OUTPUT VOLTAGE
TOLERANCE
V1: %~ -2 % (Max) 2
V2: %~ -8 % (Max) 8
V3: %~ -8 % (Max)8
I/P: 0 VAC / 264 VAC 10
O/P:FULL/ MIN LOAD
Ta:25
V1:
V2:
V3:
0.12
2.6
3.4
%~
%~
%~
-0.12
-2.6
-3.4
%
%
%
P
4
LINE REGULATION
V1: 0.5 %~ -0.5 % (Max)
V2: %~ -1 % (Max) 1
V3: %~ -1 % (Max)1
I/P: 100 VAC ~ 264 VAC
O/P:FULL LOAD
Ta:25
V1:
V2:
V3:
0
0.28
0.47
%~
%~
%~
0
-0.28
-0.47
%
%
%
P
5
LOAD REGULATION
V1: 1.5 %~ -1.5 % (Max)
V2: %~ -3 % (Max) 3
V3: %~ -3 % (Max)3
I/P: 230 VAC
O/P:FULL ~MIN LOAD
Ta:25
V1:
V2:
V3:
0
1.4
1
%~
%~
%~
0
-1.4
-1
%
%
%
P
6
CROSS REGULATION
V1: 1.5 %~ -1.5 % (Max)
V2: %~ -3 % (Max) 3
V3: %~ -3 % (Max)3
I/P: 230 VAC
O/P: Testing O/P 60%LOAD
Other O/P 40%LOAD
Change
Ta:25
V1:
V2:
V3:
0
2.2
2.7
%~
%~
%~
0
-2.2
-2.7
%
%
%
P
7
SET UP TIME
230VAC:
500
ms (Max)
I/P: 230 VAC
I/P: 115 VAC
O/P:FULL LOAD
Ta:25
230VAC/
150
ms
P
115 VAC:
500
ms (Max)
115VAC/
111
ms
8
RISE TIME
230VAC:
30
ms (Max)
I/P: 230 VAC
I/P: 115 VAC
O/P:FULL LOAD
Ta:25
230VAC/
6
ms
P
115VAC:
30
ms (Max)
115VAC/
6
ms
9
HOLD UP TIME
230VAC:
90
ms (TYP)
I/P: 230 VAC
I/P: 115 VAC
O/P:FULL LOAD
Ta:25
230VAC/
105
ms
P
115VAC:
20
ms (TYP)
115VAC/
22.4
ms
10
OVER/UNDERSHOOT TEST
+ < 5%
I/P: 230 VAC
O/P:FULL LOAD
Ta:25
TEST:
<5
%
P
11
DYNAMIC LOAD
V1:
1000
mVp-p
I/P: 230 VAC
O/P:FULL /Min LOAD
90%DUTY/1KHZ
Ta:25
150
mVp-p
P
72 Reliable Triple Output Medical Grade W
RPT- 75 series
Model: RPT- D Test Report 2 5 75 /
INPUT FUNCTION TEST
NO
TEST ITEM
SPECICATION
TEST CONDITION
RESULT
VERDICT
1
INPUT VOLTAGE RANGE
90VAC~264 VAC
I/P:TESTING
O/P:FULL LOAD
Ta:25
52V~264V
P
I/P:
LOW-LINE-3V= 87V
HIGH-LINE+ %=300 V 15
O/P:FULL/MIN LOAD
ON: 30 Sec . OFF: 30 Sec 10M IN
( AC POWER ON/OFF NO DAMAGE )
TEST: OK
2
INPUT FREQUENCY RANGE
47HZ ~63 HZ
NO DAMAGE OSC
I/P: VAC ~ 264 VAC 90
O/P:FULL~MIN LOAD
Ta:25
TEST: OK
P
3
EFFICIENCY
79% (TYP)
I/P: 230 VAC
O/P:FULL LOAD
Ta:25
79.4 %
P
4
INPUT CURRENT
230V/
1
A (TYP)
I/P: 230 VAC
I/P: 115 VAC
O/P:FULL LOAD
Ta:25
I =
0.84
A/ 230 VAC
P
115V/
1.5
A (TYP)
I =
1.42
A/ 115 VAC
5
INRUSH CURRENT
230V/
50
A (TYP)
I/P: 230 VAC
I/P: 115 VAC
O/P:FULL LOAD
Ta:25
I =
36
A/ 230 VAC
P
115V/
25
A(TYP)
I =
18
A/ 115 VAC
COLD START
6
LEAKAGE CURRENT
<
150
uA/ 2 VAC 64
for earth
leakage current
I/P: 264 VAC
O/P:Min LOAD
Ta:25
L-FG:
119
uA
P
N- : FG
119
uA
<
100
uA/2 VAC 64
for touch
leakage current
I/P: 264 VAC
O/P:Min LOAD
Ta:25
L- : V+
L-V-:
N-V+:
N-V-:
84
84
84
84
uA
uA
uA
uA
P
PROTECTION FUNCTION TEST
NO
TEST ITEM
SPECICATION
TEST CONDITION
RESULT
VERDICT
1
OVER LOAD PROTECTION
1 %~ 180 % 40
I/P: 230 VAC
I/P: 115 VAC
O/P:TESTING
Ta:25
168 %/ 230 VAC
161 %/ 115 VAC
Hiccup Mode
P
2
OVER VOLTAGE PROTECTION
CH1: 5.75V~ 6.75V
I/P: 230 VAC
I/P: 115 VAC
O/P:MIN LOAD
Ta:25
6.4 V/ 230 VAC
6.4 V/ 115 VAC
Shunt down Re- power ON
P
3
SHORT PROTECTION
SHORT EVERY OUTPUT
1 HOUR NO DAMAGE
I/P: 264 VAC
O/P:FULL LOAD
Ta:25
NO DAMAGE
Hiccup Mode
P


Produkt Specifikationer

Mærke: Mean Well
Kategori: Ikke kategoriseret
Model: RPT-75D

Har du brug for hjælp?

Hvis du har brug for hjælp til Mean Well RPT-75D stil et spørgsmål nedenfor, og andre brugere vil svare dig