Copyright © 2012–2015 Pico Technology Limited. All rights reserved.
User's Guide
ps3000d.en r2
D and D MSO model oscilloscopes
PicoScope® 3000 Series
PicoScope 3000D Series Oscilloscope & MSO User's Guide 1
Copyright © 2012–2015 Pico Technology Limited. All rights reserved. ps3000d.en r2
1 Introduction
Thank you for buying a PicoScope 3000D Series Oscilloscope from Pico
Technology.
The PicoScope 3000D Series oscilloscopes are a
range of high-specication real-time measuring
instruments that connect to the USB port of your
computer. With the PicoScope software you can
use these devices as oscilloscopes, spectrum
analyzers, and arbitrary waveform generators,
plus as a logic analyzer on the mixed-signal
models. The various model options oer
portability, deep memory, two or four analog
channels, mixed-signal inputs, fast sampling
rates and high bandwidth, making these highly
versatile oscilloscopes suitable for a wide range
of applications.
The PicoScope 3000D deep-memory oscilloscopes all feature a maximum sampling
rate of 1 GS/s, USB 3.0 connectivity, and an arbitrary waveform generator, as well as
the specications below:
Analog
channels
Digital
channels
External
trigger
Bandwidth
Buer
memory
3203D
2
-
50 MHz
64 MS
3203D MSO
16
-
3204D
2
-
70 MHz
128 MS
3204D MSO
16
-
3205D
2
-
100 MHz
256 MS
3205D MSO
16
-
3206D
2
-
200 MHz
512 MS
3206D MSO
16
-
3403D
4
-
50 MHz
64 MS
3403D MSO
16
-
3404D
4
-
70 MHz
128 MS
3404D MSO
16
-
3405D
4
-
100 MHz
256 MS
3405D MSO
16
-
3406D
4
-
200 MHz
512 MS
3406D MSO
16
-
For full specications, refer to the PicoScope 3000 Series Data Sheet on the Software
and Reference CD or to the PicoScope 3000 Series section of www.picotech.com.
Introduction2
Copyright © 2012–2015 Pico Technology Limited. All rights reserved.ps3000d.en r2
Some of the benets provided by the PicoScope 3000D Series oscilloscopes are:
Portability. Take the unit with you and simply plug it in to any Windows PC.
(Linux and Mac OS X are also supported in the Beta software).
Performance. Up to 200 MHz bandwidth, 512 MS buer, and a 1 GS/s sampling
rate.
Mixed signal capability. Display analog and digital signals on the same timebase
with the MSO models.
Flexibility. Use the device as an oscilloscope, spectrum analyzer, signal generator,
or high-speed data acquisition interface.
Programmability. The PicoScope 3000A SDK lets you write your own programs,
in your chosen programming language, to control all of the features of the scope.
Using the API functions, you can develop your own programs to collect and analyze
data from the oscilloscope. Refer to the PicoScope 3000 Series (A API)
Programmer's Guide for more information.
Long-term support. Software upgrades are available to download from
www.picotech.com. You can also call our technical specialists for support. These
services are available free of charge for the lifetime of the product.
Value for money. You don't have to pay twice for features already available on
your PC, as the PicoScope 3000 Series oscilloscope contains the specialized
hardware you need and nothing more.
Convenience. The software makes full use of the display, disk storage, user
interface, and networking built into your PC using a fast USB connection.
Five-year warranty. Your oscilloscope is covered against manufacturing faults for
ve years from the day of purchase. We don't charge a penny extra for this.
PicoScope 3000D Series Oscilloscope & MSO User's Guide 3
Copyright © 2012–2015 Pico Technology Limited. All rights reserved. ps3000d.en r2
1.1 Safety information
To prevent possible electrical shock, re, personal injury, or damage to the product,
carefully read this safety information before attempting to install or use the product.
In addition, follow all generally accepted safety practices and procedures for working
with and near electricity.
The product has been designed and tested in accordance with the European standard
publication EN 61010-1: 2010 and left the factory in a safe condition.
The following safety descriptions are found throughout this guide:
A WARNING identies conditions or practices that could result in injury or death.
A CAUTION identies conditions or practices that could result in damage to the
product or equipment to which it is connected.
Each of these safety instructions applies to all of the PicoScope 3000 Series
oscilloscopes covered by this User's Guide unless otherwise specied.
Introduction4
Copyright © 2012–2015 Pico Technology Limited. All rights reserved.ps3000d.en r2
1.1.1 Symbols
These safety and electrical symbols may appear on the product and throughout this
guide.
Symbol
Description
Direct current
Alternating current
Earth (ground) terminal
This terminal can be used to make a measurement
ground connection. It is not a safety or protective
earth.
Chassis terminal
Equipment protected throughout by double or
reinforced insulation.
Possibility of electric shock
Caution
Appearance on the product indicates a need to read
these safety instructions.
Do not dispose of this product as unsorted municipal
waste.
PicoScope 3000D Series Oscilloscope & MSO User's Guide 7
Copyright © 2012–2015 Pico Technology Limited. All rights reserved. ps3000d.en r2
1.1.5 Environment
WARNING
To prevent injury or death, do not use in wet or damp conditions, or near
explosive gas or vapor.
CAUTION
To prevent damage, always use and store your oscilloscope in appropriate
environments. Recommended temperatures and humidity conditions are
shown in the table below.
PicoScope 3000D Series
Storage
Operating
Temperature
–20 °C to +60 °C
0 °C to +40 °C
+15 °C to +30 °C
(for quoted accuracy)
Humidity
5% to 95% RH
(non-condensing)
5% to 80% RH
(non-condensing)
Altitude
2000 m
Pollution degree
2
1.1.6 Care of the instrument
The oscilloscope and its accessories contain no user-serviceable parts. Repair,
servicing, and calibration require specialized test equipment and must only be
performed by Pico Technology or an approved service provider. There may be a charge
for these services unless covered by the Pico ve year warranty.
WARNING
To prevent injury or death, do not use the instrument if it appears to be
damaged in any way, and stop use immediately if you are concerned by any
abnormal operations.
WARNING
To prevent electric shock, do not tamper with or disassemble the
oscilloscope, case parts, connectors, or accessories.
WARNING
When cleaning the oscilloscope, use a soft cloth and a solution of mild soap
or detergent in water. To prevent electric shock, do not allow liquids to enter
the casing, as this will cause damage to the electronics or insulation inside.
CAUTION
Do not tamper with or disassemble the oscilloscope, connectors or
accessories. Internal damage will aect performance.
Introduction8
Copyright © 2012–2015 Pico Technology Limited. All rights reserved.ps3000d.en r2
1.2 Conformance
1.2.1 FCC notice
This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when the equipment is
operated in a commercial environment. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference
in which case the user will be required to correct the interference at his or her own
expense.
For safety and maintenance information see section 1.1.
1.2.2 CE notice
This product meets the intent of the EMC directive 2004/108/EC and has been
tested to EN61326-1:2006 Class A Emissions and Basic Immunity.
The product also meets the intent of the Low Voltage Directive and has been
designed to meet the BS EN 61010-1:2010 (safety requirements for electrical
equipment for measurement, control, and laboratory use) standard.
PicoScope 3000D Series Oscilloscope & MSO User's Guide 9
Copyright © 2012–2015 Pico Technology Limited. All rights reserved. ps3000d.en r2
1.3 Software license conditions
The software supplied with this product is licensed, not sold. Pico Technology Limited
grants a license to the person who installs this software, subject to the conditions
listed below:
Access. The licensee agrees to allow access to this software only to persons who have
been informed of these conditions and agree to abide by them.
Usage. The software in this release is for use only with Pico Technology products or
with data collected using Pico Technology products.
Copyright. Pico Technology Limited claims the copyright of, and retains the rights to,
all material (software, documents etc.) contained in this release. You may copy and
distribute the PicoScope and PicoLog software and drivers with no modications,
additions or omissions. You may copy and modify the SDK example programs.
Liability. Pico Technology and its agents shall not be liable for any loss, damage or
injury, howsoever caused, related to the use of Pico Technology equipment or
software, unless excluded by statute.
Fitness for purpose. Because no two applications are the same, Pico Technology
cannot guarantee that its equipment or software is suitable for a given application. It
is your responsibility, therefore, to ensure that the product is suitable for your
application.
Mission-critical applications. This software is intended for use on a computer that
may be running other software products. For this reason, one of the conditions of the
license is that it excludes usage in mission-critical applications such as life-support
systems.
Viruses. This software was continuously monitored for viruses during production, but
you are responsible for virus-checking the software once it is installed.
Support. If you are dissatised with the performance of this software, please contact
our technical support sta, who will try to x the problem within a reasonable time. If
you are still dissatised, please return the product and software to your supplier within
14 days of purchase for a full refund.
Upgrades. We provide upgrades, free of charge, from www.picotech.com. We reserve
the right to charge for updates or replacements sent out on physical media.
1.4 Trademarks
Windows is a registered trademark or trademark of Microsoft Corporation, registered
in the U.S. and other countries.
Linux is the registered trademark of Linus Torvalds, registered in the U.S. and other
countries.
Mac and OS X are trademarks of Apple Inc., registered in the U.S. and other
countries.
Pico Technology Limited are trademarks of Pico Technology Limited,and PicoScope
registered in the United Kingdom and other countries.
PicoScope and Pico Technology are registered in the U.S. Patent and Trademark
Oce.
Introduction10
Copyright © 2012–2015 Pico Technology Limited. All rights reserved.ps3000d.en r2
1.5 Warranty
Pico Technology warrants upon delivery, and for a period of 5 years unless otherwise
stated from the date of delivery, that the Goods will be free from defects in material
and workmanship.
Pico Technology shall not be liable for a breach of the warranty if the defect has been
caused by fair wear and tear, wilful damage, negligence, abnormal working conditions
or failure to follow Pico Technology's spoken or written advice on the storage,
installation, commissioning, use or maintenance of the Goods or (if no advice has been
given) good trade practice; or if the Customer alters or repairs such Goods without the
written consent of Pico Technology.
PicoScope 3000D Series Oscilloscope & MSO User's Guide 11
Copyright © 2012–2015 Pico Technology Limited. All rights reserved. ps3000d.en r2
2 Product information
2.1 Connector diagrams
Standard oscilloscope connectors
The analog input channels have standard BNC connectors and standard input
impedances. They are therefore compatible with most oscilloscope probes including
x10 and switched x1/x10 variants. For optimum performance, always use the probes
supplied with your PicoScope.
Arbitrary waveform generator output
The built-in arbitrary waveform generator (AWG) can generate a waveform of almost
any shape. The generated waveform can be injected into the device you wish to test
and then analyzed as it progresses through the device to conrm correct operation, or
to highlight a fault.
·If you are using the PicoScope 6 software, refer to the PicoScope 6 User's Guide for
information on how to congure the arbitrary waveform generator.
·If you are writing your own software, refer to the PicoScope 3000 Series (A API)
Programmer's Guide.
Digital inputs for mixed-signal oscilloscopes (D MSO models only)
Alongside the analog channels, the mixed-signal 3000D MSO Series oscilloscopes also
feature 16 digital inputs. The PicoScope software allows you to view both digital and
analog signals simultaneously. Your digital inputs are easily manageable, and can be
reordered, grouped, and renamed.
Features such as triggering and serial decoding can also be used on the digital
channels.
See Digital inputs for PicoScope 3000D Series MSO models for more information.
External trigger (EXT) input (D models only)
The EXT Trigger input can be used as a trigger source. Select it from the menu in the
PicoScope software, or by using a function call if you are writing your own software.
The EXT input uses dedicated circuitry, in combination with a software-congurable
threshold, to detect a trigger signal. This leaves the analog channels free for viewing
signals. The input characteristics of the EXT input are matched to a scope channel so
that the supplied, compensated probes can be used with the EXT input to give high
vertical accuracy.
Where trigger timing accuracy is critical, we recommend using one of the main input
channels as the trigger source. These channels use accurate digital triggering (to one
sample period) and have a vertical resolution of 1 LSB.
USB port
All models have a SuperSpeed USB 3.0 port for connection to a PC. See Connectivity,
power and installation for more information.
Earth terminal
External noise may sometimes interfere with your measurements if the PicoScope is
used with a computer which does not have a ground connection. If this is the case,
connect the earth terminal to an external ground point (for example, on the system
you are testing) to provide a ground reference for the oscilloscope.
Product information12
Copyright © 2012–2015 Pico Technology Limited. All rights reserved.ps3000d.en r2
2.1.1 2-channel model connector diagram
PicoScope 3203D, 3204D, 3205D, and 3206D.
A. Input channel A
B. Input channel B
C. Probe compensation pin (above)
LED (below)
Flashes green when the oscilloscope is sampling data
D. External trigger (EXT) input
E. Arbitrary waveform generator output
F. Earth terminal
G. USB port
Product information14
Copyright © 2012–2015 Pico Technology Limited. All rights reserved.ps3000d.en r2
2.1.3 2-channel MSO model connector diagram
PicoScope 3203D MSO, 3204D MSO, 3205D MSO, and 3206D MSO.
A. Input channel A
B. Input channel B
C. Probe compensation pin (above)
LED (below)
Flashes green when the oscilloscope is sampling data
D. Digital inputs D0 to D15
E. Arbitrary waveform generator output
F. Earth terminal
G. USB port
PicoScope 3000D Series Oscilloscope & MSO User's Guide 15
Copyright © 2012–2015 Pico Technology Limited. All rights reserved. ps3000d.en r2
2.1.4 4-channel MSO model connector diagram
PicoScope 3403D MSO, 3404D MSO, 3405D MSO, and 3406D MSO.
A. Input channel A
B. Input channel B
C. Input channel C
D. Input channel D
E. Probe compensation pin (above)
LED (below)
Flashes green when the oscilloscope is sampling data
F. Digital inputs D0 to D15
G. Arbitrary waveform generator output
H. Earth terminal
I. USB port
J. DC power input
PicoScope 3000D Series Oscilloscope & MSO User's Guide 17
Copyright © 2012–2015 Pico Technology Limited. All rights reserved. ps3000d.en r2
2.2 Connectivity, power and installation
1. PicoScope software installation
Before setting up your PicoScope 3000 Series oscilloscope, it is recommended that you
rst install the PicoScope 6 software by following the instructions in the supplied Quick
Start Guide.
There are dierent connectivity and power supply options for each oscilloscope model
depending on its specications.
2. USB connectivity
For best results, use only the high-quality USB cables supplied with your PicoScope
·If your computer has a USB 3.0 port, connect the oscilloscope to it using the USB
3.0 cable supplied (image below).
·The oscilloscope is also compatible with a single USB 2.0 port, and can be
connected using the USB 3.0 cable (image below), provided the port supplies the
required power. The PicoScope software will notify you if the port is not suitable.
·If your computer does not have a USB 3.0 port or a suitable USB 2.0 port
available, you can use a double-headed USB 2.0 cable (available separately)
instead, which allows you to use a second USB port for additional power.
Connector A : Connect to the USB port on
your PC
Connector B : Connect to the USB port on
your PicoScope 3000 Series oscilloscope
PicoScope 3000D Series Oscilloscope & MSO User's Guide 19
Copyright © 2012–2015 Pico Technology Limited. All rights reserved. ps3000d.en r2
2.3 Minimum system requirements
To ensure that your PicoScope 3000 Series oscilloscope operates correctly, you must
have a computer with the system requirements and one of the operating systems
shown in the table below. The performance of the oscilloscope will be better with a
more powerful PC, and will greatly benet from a multi-core processor.
Item
Specication
Operating system
Windows 7, Windows 8 (not Windows RT), Windows 10*.
32 bit and 64 bit versions.
Beta software is also available for Linux and OS X operating
systems.
Processor
As required by the operating system
Memory
Free disk space
Ports
USB 3.0 USB 2.0 or port(s)**
* PicoScope version 6.11 and SDKs are compatible with Windows XP SP3 and
Vista SP2 in addition to the Windows versions listed above. For best
performance we recommend Windows 7 or later.
** See Connectivity, power and installation for more information.
Product information20
Copyright © 2012–2015 Pico Technology Limited. All rights reserved.ps3000d.en r2
2.4 Pack contents
All PicoScope 3000 Series oscilloscope kits contain:
·PicoScope 3000 Series oscilloscope
·Quick Start Guide
·Software and reference CD
·USB 3.0 cable*
Each model is also supplied with probes and additional items as shown below.
Probes
(x1/x10, 1.2 m)
Digital cable
Test clips
AC power
adaptor
3203D
60 MHz (x2)
-
-
-
3203D MSO
3204D
150 MHz (x2)
-
-
3204D MSO
3205D
150 MHz (x2)
-
-
3205D MSO
3206D
250 MHz (x2)
-
-
3206D MSO
3403D
60 MHz (x4)
-
-
3403D MSO
3404D
150 MHz (x4)
-
-
3404D MSO
3405D
150 MHz (x4)
-
-
3405D MSO
3406D
250 MHz (x4)
-
-
3406D MSO
* .For more information on USB cables, see Connectivity, power and installation
Each PicoScope 3000 Series oscilloscope is supplied with two or four probes that have
been selected for use with that model. For optimal performance, always use the probes
provided. Although other oscilloscope probes can be used, the specied performance
cannot be guaranteed. Replacement probes can be ordered at www.picotech.com.