Cleaning
Maintenance
Troubleshooting
Specification 
1.  Please clean the surface of the device before using. 
Wipe the device with medical alcohol (70% isopro-
pyl alcohol)first, and then let it dry in air or clean it 
by dry clean fabric. When cleaning the device with 
water, the water temperature should be lower than 
60°C
2.  Using the medical alcohol to disinfect the product 
after use, prevent from cross infection for next time 
use.
3.  The best storage environment of the device is - 25°C 
to 70°C ambient temperature and not higher than 
90% relative humidity. 
Note: 
1.  Do not sterilize, autoclave or immerse this device 
in liquid. Do not pour or spray any liquids onto the 
device.
2.  Do not use caustic or abrasive cleaning agents, or 
any cleaning agent containing ammonium chloride 
or isopropyl alcohol.
Recommends user to return this device to the man-
ufacturer perform the following checks every 24 
months.
-  Inspect the equipment for mechanical and functional 
damage or deterioration.
-  Ensure all user interface keys and accessories function 
normally.
Note:  Manufacturer use Index2 SpO2 simulator to ver-
ify operation of the pulse oximeter equipment.
Symptoms Check points Corrections
SpO2 or pulse 
rate cannot 
displayed
Applied nger 
improperly.
Place the nger prop-
erly and try again.
SpO2 is too low to 
detect
Try again; go to consult 
with your physician if 
you are sure the device 
works well.
SpO2 or pulse 
rate are not dis-
played stably
Applied nger 
improperly.
Place the nger prop-
erly and try again.
Finger is shaking or 
body is moving.
Keep body steady
No display 
when button is 
pressed
Batteries run down
Replace with new 
batteries
Batteries not 
inserted correctly.
Re-insert batteries
The display 
disappears 
suddenly
The device will auto 
power o when it 
gets no signal.
Normal
Low battery
Replace with new 
batteries
SpO
2
 
Measuring range 35%~99%, (the resolution is 1%)
Accuracy 70%~99%: ±2%, Below 35~69%: unspecied
Optical Sensor The wavelength of red LED is 660 nm and 
Infrared LED is 905/880 nm with maximum 
optical output power of 4 mW/sr.
Pulse 
Measuring range 30 bpm~250 bpm (the resolution is 1 bpm)
Accuracy ±3 bpm
Power source AAA × 2 (Alkaline)
Battery life Continually for 16 hours with two alkaline 
batteries
Operating 
Condition
Temperature: 5°C~40°C (41°F ~ 104°F), 
Relative Humidity: 15-95% (non condensing), 
Atmospheric pressure: 700hPa ~ 1060hPa, 
Attitude: -1,280 to 12,000 feet (-390m to 3,658m)
Storage / 
Transportation 
Condition
Temperature: -25°C~+70°C(-13°F ~ 158°F), 
Relative humidity: 15-90%(non condensing), 
Atmospheric pressure: 700hPa ~ 1060hPa, 
Attitude: -1,280 to 12,000 feet (-390m to 3,658m)
Dimensions 63.5(L) × 34(W) × 35(H) mm
Weight About 37g (without the batteries)
Standards IEC60601-1-2, Class B, IEC60601-1, Type BF, 
ISO80601-2-61, IEC60601-1-11
Type BF applied parts
IP Classication IP22: Protection against harmful ingress of water 
and particulate matter
Attention
Serial number
Manufacturer
Date of manufacture
Authorized representative in the European 
Community
Note:
-  A description of the effect on displayed and transmitted 
SpO2 and pulse rate:
-  Data averaging: 4 seconds for SpO2; 8 seconds for pulse 
rate.
- Data update delay: Less than 2 seconds.
-  The oximeter is calibrated in the factory before sale. 
There is no need to calibrate during its life cycle.
 Note:  If the unit does not work, return it to your 
dealer. Under no circumstance should you 
disassemble and repair the unit by yourself.
Model: SB220
 www.rossmax.com
 Fingertip Pulse Oximeter
Rossmax InnoTek Corp.  
12F., No. 189, Kang Chien Rd., Taipei, 
114, Taiwan.
CMC Medical Devices & Drugs S.L.
C/ Horacio Lengo Nº 18, CP 29006, 
Málaga, Spain
   WARNING: The symbol on this product means that it's an electronic product 
and following the European directive 2012/19/EU the electronic products have 
to be dispose on your local recycling centre for safe treatment.
IN0SB221000000013
Warranty Card
This instrument is covered by a 2 year guarantee from the date of purchase, batteries 
and accessories are not included. The guarantee is valid only on presentation of the 
guarantee card completed by the dealer confirming date of purchase or the receipt. 
Opening or altering the instrument invalidates the guarantee. The guarantee does 
not cover damage, accidents or non-compliance with the  instruction manual. 
Please contact your local seller/dealer or www.rossmax.com.
Customer Name:  __________________________________
Address:  _______________________________________
Telephone:  _____________________________________
E-mail address:  ___________________________________
Product Information:  ______________________________
Date of purchase: _________________________________
Store where purchased: 
______________________________________________
RI_IB_SB220_EN_ 
2Yr_TP_ver2128
 Note: 
1.  The product is with a service life of 5 years in the course 
of regular use.
2.  If any serious incident (e.g. death) has occurred in 
relation to the device should be reported to the 
dealer, manufacturer, and the competent authority of 
the Member State in which the user and/or patient is 
established.
3. The text is subject to change without further notice.
EMC guidance and manufacturer’s declaration
Recommended separation distances between portable and mobile RF communications equipment and 
the ME equipment
The Finger-tip pulse oximeter is intended for use in an electromagnetic environment in which radiated RF 
disturbances are controlled. The customer or the user of the Finger-tip pulse oximeter can help prevent elec-
tromagnetic interference by maintaining a minimum distance between portable and mobile RF communi-
cations equipment (transmitters) and the Finger-tip pulse oximeter as recommended below, according to 
the maximum output power of the communications equipment.
Rated maximum output 
power of transmitter / W
Separation distance according to frequency of transmitter / m
150 kHz to 80 MHz , 
d=[3.5/V1]√P
80 MHz to 800 MHz , 
d=[3.5/E1]√P
800 MHz to 2,5 GHz , 
d=[3.5/E1]√P
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.7 3.7 7.37
100 11.67 11.67 23.33
Declaration – electromagnetic emissions and immunity – for EQUIPMENT and SYSTEMS that are not LIFE-
SUPPORTING and are specied for use only in a shielded location
The Finger-tip pulse oximeter declaration – electromagnetic immunity
The Finger-tip pulse oximeter system is intended for use in the electromagnetic environment specied 
below. The customer or the user of the Finger-tip pulse oximeter system should assure that it is used in 
such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3V
Portable and mobile RF communications 
equipment should be used no closer to any part 
of the EQUIPMENT or SYSTEM including cables, 
than the recommended separation distance 
calculated from the equation applicable to the 
frequency of the transmitter. Interference may 
occur in the vicinity of equipment marked with 
the following symbol.  
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3V/m
Declaration – electromagnetic immunity
The Finger-tip pulse oximeter system is intended for use in the electromagnetic environment specied 
below. The customer or the user of the Finger-tip pulse oximeter system should assure that it is used in 
such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment-
guidance
Electrostatic 
discharge (ESD) 
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete 
or ceramic tile. If oors are covered 
with synthetic material, the 
relative humidity should be at 
least 30 %.
Electrical fast 
transient/burst 
IEC 61000-4-4
±2 kV for power sup-
ply lines
±1 kV for input/
output lines
±2 kV for power sup-
ply lines
Mains power quality should be 
that of a typical commercial or 
hospital environment.
Surge IEC 
61000-4-5
± 1kV dierential mode
± 2kV common mode
± 1kV dierential mode
± 2kV common mode
Mains power quality should be 
that of a typical commercial or 
hospital environment.
Voltage dips, 
short interrup-
tions and voltage 
variations on 
power supply 
input lines IEC 
61000-4-11
<5% UT(>95% dip in 
UT) for 0,5 cycle
40% UT(60% dip in UT) 
for 5 cycles
70% UT(30% dip in UT) 
for 25 cycles
<5% UT(>95% dip in 
UT) for 5 s
<5% UT(>95% dip in 
UT) for 0,5 cycle
40% UT(60% dip in UT) 
for 5 cycles
70% UT(30% dip in UT) 
for 25 cycles
<5% UT(>95% dip in 
UT) for 5 s
Mains power quality should be 
that of a typical commercial or 
hospital environment. If the user 
of the EQUIPMENT or SYSTEM 
requires continued operation 
during power mains interruptions, 
it is recommended that the EQUIP-
MENT or SYSTEM be powered 
from an uninterruptible power 
supply or a battery. 
Power frequency 
(50/60 Hz) 
magnetic eld 
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic elds 
should be at levels characteristic of 
a typical location in a typical com-
mercial or hospital environment.
Declaration – electromagnetic emissions
The Finger-tip pulse oximeter is intended for use in the electromagnetic environment specied below. The 
customer or the user of the Finger-tip pulse oximeter should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment-guidance
CE emissions CISPR11 Group 1 The Finger-tip pulse oximeter uses RF energy only for 
its internal function. Therefore, its RF emissions are 
very low and are not likely to cause any interference 
in nearby electronic equipment.
RF emissions CISPR 11 Class B The Finger-tip pulse oximeter is suitable for use in all 
establishments, including domestic establishments 
and those directly connected to the public low-voltage 
power supply network that supplies buildings used for 
domestic purposes.
Harmonic emissions IEC 61000-3-2 Class A
Voltage uctuations/icker emis-
sions IEC 61000-3-3
Complies