15. Technical Support and Warranty
12. Maintenance & tips
Make sure the sensor and probe cavity is clean otherwise it will affect accuracy.
Cleaning method for probe:
1.Use the cotton stick or soft cloth with water or alcohol to wipe the casing.
2.Use the cotton stick or soft cloth with alcohol to wipe the sensor surface or probe
cavity gently. Don't use thermometer before alcohol is vaporized.
Read this manual book thoroughly before use. Make sure battery is well installed.
It is not allowed to put the thermometer in any liquid or expose to strong sunlight
or extremely low temperature.
Strong crash or hit to the product will cause its damage.
Do not dismantle this thermometer by yourself.
Open the battery lid and take out exhausted battery.
Put into 2 AAA batteries and close up battery lid. After new battery is installed,
the color of backlight turns in the order of green , orange and red , each of them
flashing one time respectively, with a “beep” heard. If no beep is heard, please check
Notice:
1.Take out battery in case the thermometer is not used for a long time. Don't put the
battery to fire.
2.Dispose battery according to local regulations.
11. Battery replacement
Keep the thermometer from children's reach.
Do not use the thermometer under circumstance of strong electromagnetic interfere.
(
fig.11.1
)
Standard list
Medical device software - Software life-cycle processes
ISO 80601-2-56
Medical electrical equipment part 2-56: particular
requirements for basic safety and essential performance
of clinical thermometer for body temperature measurement
Disposal
NOTES:
Please act according to the native law to proceed to handle the battery and wastes.
Take out the battery if you are not going to use the unit for a long time.
IEC 62366 Medical devices – Application of usability engineering to
medical devices
Biological evaluation of medical devices - Part 1:
Evaluation and testing within a risk management process
IEC 15223-1
Symbols for use in the labeling of medical devices
IEC 60601-1-11
Medical electrical equipment – Part 1-11: General requirements
for basic safety and essential performance – Collateral standard:
Requirements for medical electrical equipment and medical
electrical systems used in home healthcare environment
IEC 60601-1-2
Medical electrical equipment -- Part 1-2: General requirements
for basic safety and essential performance - Collateral standard:
Electromagnetic compatibility - Requirements and tests
IEC 60601-1 Medical electrical equipment Part 1: General requirements
for basic safety and essential performance
Information supplied by the manufacturer with medical devices
Standard Specification for Infrared Thermometer for
Intermittent Determination of Patient Temperature
IEC 60601-1-6
Medical electrical equipment – Part1-6: General requirements
for basic safety and essential performance – Collateral
standard: Usability
Dispose of the device in accordance with the regulation applicable at the place
of operation. Dispose of at public collection point in the EU countries –
WEEE Directive.
If you have any queries, please refer to the local authorities responsible for
waste disposal.
Dispose of empty battery at your retail store or at appropriate collection sites
according to national or local regulations to protect environment.
Read instruction manual before use
Type BF applied part
Batch
Serial number
LOT
SN
Manufacturer information
IP code of the device: this device’s grade of against ingress of solid
IP22
parliament and of the council of 8 June 2011
RoHS
COMPL IANT
Follow operating instructions
Guidance and manufacturer´s declaration – electromagnetic emission –
for all EQUIPMENT AND SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic emission
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage fluctuations
flicker emissions
IEC 61000-3-3
Class B
N/A
N/A
Emissions test
Group 1
Electromagnetic environment - guidance
The Infrared Thermometer uses RF energy
only for its internal function. There for, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
The Infrared Thermometer suitable for use
in all establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
The infrared thermometer is intended for use in the electromagnetic environment
specified below.
The customer or the user of infrared thermometer should assure that it is used in
such an environment..
The infrared thermometer is intended for use in the electromagnetic environment
specified below.
The customer or the user of infrared thermometer should assure that it is used in
such an environment.
Guidance and manufacturer's declaration – electromagnetic immunity –
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic emission
Guidance and manufacturer´s declaration – electromagnetic emission –
for all EQUIPMENT AND SYSTEMS
NOTE UT is the a. c. mains voltage prior to application of the test level.
Guidance and manufacturer´s declaration – electromagnetic immunity
Immunity test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV,
± 15 kV air
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV,
± 15 kV air
Floors should be wood,
concrete or
ceramic tile. If floors are
covered with
synthetic material, the
relative humidity
should be at least 30 %.
Compliance level
IEC 60601
test level
Electromagnetic
environment-guidance
Electrostatic
transient /
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Voltage dips,
short
interruptions
and
voltage
variations
on power
supply
input lines
IEC
61000-4-11
± 2 kV for
power
supply lines
± 1 kV for
input/output
lines
± 1 kV
differential
mode
± 2 kV
common mode
0 % UT; 0,5
cycle g) At 0°,
45°, 90°, 135°,
180°, 225°,
270° and 315°
0 % UT; 1
cycle and 70 %
UT; 25/30
cycles Single
phase: at 0°
0 % UT;
250/300 cycle
N/A
N/A
30 A/m 30 A/m
N/A
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of the Infrared
thermometer requires
continued operation
during power mains
interruptions, it is
recommended that the
Infrared thermometer be
powered from an
uninterruptible power
supply or a battery.
The infrared thermometer is intended for use in the electromagnetic environment
specified below.
The customer or the user of infrared thermometer should assure that it is used in
such an environment.
For transmitters rated at a maximum output power not listed above the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
Rated
maximum
output of
transmitter
W
150 kHz to 80 MHz
outside ISM and
amateur radio bands
150 kHz to 80 MHz
in ISM and amateur
radio bands
180 MHz to 800
MHz
800 MHz to 2.7 GHz
Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM -for
EQUIPMENT and SYSTEMS
Separation distance according to frequency of transmitter
m
Recommended separation distances between
portable and mobile RF communications equipment and the Infrared thermometer
The Infrared thermometer is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the
Infrared thermometer can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment
(transmitters) and the Infrared thermometer as recommended below, according to
the maximum output power of the communications equipment
b. Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in
the location in which the Infrared thermometer is used exceeds the applicable RF
compliance level above, the Infrared thermometer should be observed to verify
normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the Infrared thermometer.
c. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than
3V/m.
d=[ ] p
V1
3.5 d=[ ] p
V
2
12
0.01
1
0.1
10
100
0.12
1.2
0.38
3.8
12
0.20
2.00
0.63
6.32
20.00
0.035
0.35
0.11
1.10
35
0.07
0.70
0.22
2.21
70
NOTE 1 At 80 MHz and 800 MHz, the higher frequency ra nge applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is
affected by absorption and reflection from structures, objects and people.
a. The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz
are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283
MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz
and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7
MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17
MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and
50,0 MHz to 54,0 MHz.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
6 V in ISM and
amateur radio
bands between
0,15 MHz and
80 MHz
10 V/m
80 MHz to 2.7
GHz
385MHz-5785M
Hz Test
specifications for
ENCLOSURE
PORT
IMMUNITY to
RF wireless
communication
equipment
(Refer to table 9
of IEC
60601-1-2:2014)
N/A
10 V/m
80 MHz to 2.7 GHz
385MHz-5785MHz
Test specifications
for ENCLOSURE
PORT IMMUNITY to
RF wireless
communication
equipment (Refer to
table 9 of IEC
60601-1-2:2014)
80 MHz to 800 MHz
800 MHz to 2.7 GHz
Portable and mobile RF
communications equipment should
be used no closer to any part of the
Infrared thermometer, including
cables, than the recommended
separation distance calculated
from the equation applicable to the
frequency of the transmitter.
where p is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in metres (m).b
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the
vicinity of equipment marked with
the following symbol:
Recommended separation distance
Immunity test Compliance level
IEC 60601
test level
Electromagnetic
environment-guidance
d=[ ] p
V
1
3.5
d=[ ] p
V2
12
d=[ ] p
E1
3.5
d=[ ] p
E1
7
Normalized symbols
1) Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify that they are
operating normally.
2) Use of accessories, transducers and cables other than those specified or provided
by the manufacturer of this the Infrared Thermometer could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this equipment
and result in improper operation.
3) Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any
part of the Infrared Thermometer, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result.
If you need further assistance, please contact the Aluratek support department
for troubleshooting prior to returning this device.
E-mail: support@aluratek.com
Web: www.aluratek.com/helpdesk
Local (Irvine, CA): 714-586-8730
Toll free: 1-866-580-1978
Aluratek warrants this product against defect in material or workmanship for
1 YEAR from the date of purchase.
For more information, please visit: https://aluratek.com/warranty-return-policy
You can register your product online at: https://aluratek.com/product-registration