Læs gratis den danske manual til Mean Well RPD-60A (4 sider) i kategorien ikke klassificeret. Denne vejledning er vurderet som hjælpsom af 40 personer og har en gennemsnitlig bedømmelse på 4.8 stjerner ud af 20.5 anmeldelser. Har du et spørgsmål om Mean Well RPD-60A, eller vil du spørge andre brugere om produktet?

Side 1/4
Installation Manual
Type
Medical Switching Power Supply
(1) Enclosed (Families: MSP)
(2) Modular (Families: MP, NMP)
(3) PCB (Families: MPS, MPD, MPT, MQ, RPS, RPD, RPT, RPTG, NEM, PM)
(4) Adaptor (Families: GSM, MES)
Installation
(1) Enclosed Type
(a) Before any installation or maintenance work, please disconnect your system from the utility.
Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient
temperature may increase the internal component temperature and will require a de-rating in
output current. Please refer to the specification sheets to receive the optimum mounting
position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance
must be kept when the adjacent device is a heat source.
(d) Keep enough insulation distance between mounting screws and internal components of
Enclosed units. Please refer to case drawing on specifications to receive the maximum length of
mounting screw.
(e) Recommended wires of Enclosed units are shown as below.
AWG 18 16 14 12 10 8
Rated Current of
Equipment (Amp) 6A 6-10A 10-16A 16-25A 25-32A 32-40A
Cross-section of
Lead(mm2 ) 0.75 1.00 1.5 2.5 4 6
Note: The suggested maximum current above only suit for 1-4 strips, 5 or more
strips should not exceed 80% of current rating.
Make sure that all strands of each stranded wire enter the terminal connection and the screw
terminals are securely fixed to prevent poor contact.
(f) For other information about the products, please refer to www.meanwell.com for details.
(2) Modular Type
(a) Before any installation or maintenance work, please disconnect your system from the utility.
Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient
temperature may increase the internal component temperature and will require a de-rating in
output current. Please refer to the specification sheets to receive the optimum mounting
position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance
must be kept when the adjacent device is a heat source.
(d) Keep enough insulation distance between mounting screws and internal components of
Modular units. Please refer to case drawing on specifications to receive the maximum length of
mounting screw.
(e) Recommended wires of Modular units are shown as below.
AWG 18 16 14 12 10 8
Rated Current of
Equipment (Amp) 6A 6-10A 10-16A 16-25A 25-32A 32-40A
Cross-section of
Lead(mm2 ) 0.75 1.00 1.5 2.5 4 6
Note: The suggested maximum current above only suit for 1-4 strips, 5 or more
strips should not exceed 80% of current rating.
Make sure that all strands of each stranded wire enter the terminal connection and the screw
terminals are securely fixed to prevent poor contact.
(f) For other information about the products, please refer to www.meanwell.com for details.


Produkt Specifikationer

Mærke: Mean Well
Kategori: ikke klassificeret
Model: RPD-60A

Har du brug for hjælp?

Hvis du har brug for hjælp til Mean Well RPD-60A stil et spørgsmål nedenfor, og andre brugere vil svare dig




ikke klassificeret Mean Well Manualer

ikke klassificeret Manualer

Nyeste ikke klassificeret Manualer

Lindy

Lindy 42661 Manual

21 August 2024
LG

LG WD-1485RD Manual

21 August 2024
ATen

ATen VS1808T 8 Manual

19 August 2024
Dormakaba

Dormakaba X-10 Manual

19 August 2024
Lifan

Lifan LF-160F-2 Manual

19 August 2024
Martin

Martin Mach CW215 Manual

18 August 2024
Microchip

Microchip MCP2551 Manual

16 August 2024