PCE Instruments PCE-CPC 100 Manual

PCE Instruments Måling PCE-CPC 100

Læs gratis den danske manual til PCE Instruments PCE-CPC 100 (15 sider) i kategorien Måling. Denne vejledning er vurderet som hjælpsom af 3 personer og har en gennemsnitlig bedømmelse på 4.0 stjerner ud af 2 anmeldelser. Har du et spørgsmål om PCE Instruments PCE-CPC 100, eller vil du spørge andre brugere om produktet?

Side 1/15
PCE- CPC 100
Dust Meter
2
2
2
22 27
27
27
2727/
PCE-CPC 100 Particle Counter
Description
The PCE-CPC 100 online particle counter adopts the principle of optical scattering, which can accurately detect and
calculate the number of suspended particles of different particle sizes in the air per unit volume. It can output the
particle count of 5 channels of 0.3μm, 0.5μm, 1.0μm, 5.0μm and 10μm at the same time (default unit is pcs/28.3L,
available to switch unit to be pcs/m³).
Features
l- Rea time output particle quantities of 0.3μm, 0.5μm, 1.0μm, 5.0μm, 10μm in pcs/28.3L or pcs/m³.
Sound and light alarm once particle quantity exceeds the set threshold.
ModBus communication protocols available.
- - Real time display cleanroom ISO 14644 1 grade level.
Output units switchable between pcs/28.3L and pcs/m³.
Constant flow gas sampling system to ensure stable sampling.
Industrial grade laser for high reliability.
Working Principle
Air sampling is carried out by a fan. When the particles in the sampled gas pass through a light source (laser) and
other beams, light scattering occurs; the scattered light is converted into an electrical signal (pulse) through a
photoelectric converter, and the larger the particle, the pulse signal is obtained. The larger the wave value (wave
peak value), the number of particles with different particle sizes can be obtained through the wave peak value and
the number of pulses at this time.
According to the block diagram , the light source part o consists of a laser tube that emits light to me f PCE-CPC 100
detect particles and a drive circuit, the detection part consists of a photosensitive element that receives reflected
light and an amplifier circuit, and the data processing and communication output are completed by a
microprocessor.
The particle detection of PCE CPC 100 is the gas flow generated by the operation of the fan, and the particles pass -
through the detection chamber. The light from the laser tube will be scattered by the particles and converted into
electrical signals by the photosensitive device. After the electrical signal is processed by amplifying circuit, filtering
and MCU, it will be converted into digital signal output.


Produkt Specifikationer

Mærke: PCE Instruments
Kategori: Måling
Model: PCE-CPC 100

Har du brug for hjælp?

Hvis du har brug for hjælp til PCE Instruments PCE-CPC 100 stil et spørgsmål nedenfor, og andre brugere vil svare dig