Læs gratis den danske manual til PCE Instruments PCE-RD 75 (7 sider) i kategorien Måleudstyr. Denne vejledning er vurderet som hjælpsom af 10 personer og har en gennemsnitlig bedømmelse på 4.8 stjerner ud af 5.5 anmeldelser. Har du et spørgsmål om PCE Instruments PCE-RD 75, eller vil du spørge andre brugere om produktet?

Side 1/7
User manuals in various languages (français, italiano, español, português, nederlands,
türk, polski) can be found via our product search on:
www.pce-instruments.com
PCE-RD 75
USER MANUAL
ENGLISH
RADON DETECTOR
2
PCE-RD 75 radon detector is a high precision radon concentration monitoring and alarm instrument, used
don pulse ionization chamber sensors. Historical data can be viewed locally, or the device can be connected
to the computer to view the PDF data report in the device USB drive. Our radon detectors communicate
with your computer via the USB ash drive protocol, so you don‘t need to install any additional software
on your computer. Use 2.1 inch high resolution capacitive touch color screen, UI interface is easy to operate.
When the detection value exceeds the alarm threshold, an alarm can be issued to remind the user to
ventilate in time.
The concentration limit for radon in the civil built environment varies from country to country (our equip-
ment defaults to 148Bq/m³), but the following basic standards are generally used.
The simplest way to reduce indoor radon concentration is to increase indoor ventilation. When the radon
concentration exceeds 4pCi/L(148 Bq/m³), ventilate for at least 10 minutes or more.
»Used for environmental radon monitoring in mines, basements and poorly ventilated residential
buildings.
»Investigation and evaluation of radon concentration in soil at construction site.
»It can be used for mineral resource exploration, engineering geological exploration, environmental radon
concentration monitoring in related departments of nuclear industry, and radiation detection of radon
concentration in geological investigation.
1.PRODUCT INTRODUCTION
3.LIMITS ON RADON CONCENTRATION
2.WHERE IT CAN BE USED
China standards ≤ 150Bq/m³
United States (EPA) standards ≤ 4pCi/L
( Or ≤ 148Bq/m³)
(WHO) recommended standards -≤ 100Bq/m³ , Corrective action is strongly recom
mended at levels greater than 300Bq/m³
European (EURATOM) standards ≤ 300Bq/m³