HEATED HUMIDIFIER
ClimateLine / ClimateLineMAX ™
HEATED AIR TUBING
Information Guide
English • Svenska • Dansk • Norsk • Suomi • Eesti简体中文繁體中文
Indications for use
The H5i is indicated for the humidification of the air delivered from a CPAP or bilevel device. The H5i is
for use only as recommended by a physician. The H5i is intended for single patient re-use in the home
environment and re-use in a hospital/institutional environment.
Contraindications
The H5i is contraindicated for use with patients whose upper (supraglottic) airway has been bypassed.
Please refer to the S9 Series Information Guide for contraindications associated with CPAP or bilevel
therapy.
Troubleshooting
If there is a problem, try the following suggestions. If the problem cannot be solved, contact your
equipment supplier or ResMed. Do not attempt to open these devices.
English
Problem/possible cause Solution
No display
Power is not connected. Ensure the power cable is connected and the power outlet
(if available) is on.
The DC plug is partially inserted into the
back of the device.
Fully insert the DC plug.
The S9 and H5i are not connected
correctly.
Ensure that the S9 and H5i are securely attached.
Insufficient air delivered from the device
The S9 is not working correctly. See the Troubleshooting section of the S9 Series
Information Guide.
English
2
Air tubing is not connected properly. Check air tubing.
Air tubing is blocked, pinched or
punctured.
Unblock or free the air tubing. Check the air tubing for
punctures.
The H5i flip lid is not latched correctly. Close the flip lid ensuring that it clicks into place.
The H5i flip lid seal is not fitted correctly. Make sure the flip lid seal is facing the right way up and
fitted securely.
Incorrect air tubing selected. If you are using the SlimLine or Standard air tubing ensure
that you have the correct air tubing selected via the menu.
Displays error message: Heated tube fault, replace tube
Device has been left in a hot
environment.
Allow to cool before re-use. Disconnect the power cord
and then reconnect it to restart the device.
There is a fault in your ClimateLine or
ClimateLineMAX air tubing.
Discontinue using your ClimateLine or ClimateLineMAX air
tubing and contact your clinician/service provider. Use
Standard or SlimLine tubing in the interim.
Displays error message: Humidifier fault, replace humidifier
There is a fault in your H5i heated
humidifier.
Discontinue using your H5i and contact your clinician/
service provider.
Device has been left in a hot
environment.
Allow to cool before re-use. Disconnect the power cord
and then reconnect it to restart the device.
Refilling the humidifier with cold water
while humidifier is still hot after therapy.
Allow the humidifier to cool before re-filling.
Filling the humidifier with ice cold water
on a warm day or with hot water.
Use room temperature water.
Displays error message: Tube blocked, please check your tube
Air tubing is blocked. Check your air tubing and remove any blockages.
Displays error message: High leak, please check system setup and all connections
There is excessive leak. Adjust position of mask and headgear.
Air tubing is not connected properly. Connect firmly at both
ends.
The H5i flip lid is not latched correctly. Close the flip lid ensuring that it clicks into place.
Problem/possible cause Solution