
This  instrument  has  been  designed,  manufactured  and  tested 
according to IEC 61010: Safety requirements for Electronic measuring 
apparatus,  and  delivered  in  the  best  condition  after  passed  the 
inspection. This instruction manual contains warnings and safety rules 
which  must  be  observed  by  the  user  to  ensure  safe operation  of  the 
instrument  and  retain  it  in  safe  condition. Therefor
these operating instructions before using the instrument.
●  Read  through  and  understand  the  instructions  contained  in  this 
manual before starting to use the instrument.
●  Keep  the  manual  at  hand  to  enable  quick  reference  whenever 
● The instrument is to be used only in its intended applications.
●  Understand and follow  all the safety instructions contained  in the 
It  is essential  that  the  above  instructions are  adhered  to.  Failur
follow  the above  instructions  may impair  the  protection  provided by 
the  instrument,  and  may  cause  injury,  instrument  damage  and/or 
damage to the equipment under test.
The  symbol    indicated  on  the  instrument  means  that  the  user #
must refer  to the related  parts in the  manual for safe  operation of the 
instrument. It is essential to read the instructions wherever the symbol  
#  DANGER is reserved  for conditions and  actions that are likely  to 
cause serious or fatal injury.
#  WARNING is  reserved for  conditions and actions  that can  cause 
#  CAUTION  is  reserved  for conditions  and  actions  that  can  cause 
injury or instrument damage.
● Never make measurement on a circuit in the following categories;
  Measurement category IV(CAT IV):over 600V
  Measurement category III(CAT III):over 1000V
●  Do  not  att empt  to  make  me asurement  in  the  pr esence  of 
flammable    gasses.  Otherwise,  the  use  of  the  instrument  may 
cause sparking, which can lead to an explosion.
Never attempt to use the instrument if its surface or your hand is wet.
Do not exceed the maximum allowable input of any measuring ranges.
Never open the Battery compartment cover during a measurement.
To avoid getting electrical shock by touching the equipment under 
test or its surroundings, be sure to wear insulated protective gears. 
●  Never  attempt  to make  any  measurements  if  the instrument  has 
any structural abnormalities, such as a crack, or if the cover 
●  The  instrument  should  be  used  only  in  its  intended  applications 
or  conditions.  Otherwise,  safety  functions  equipped  with  the 
instrument  do  not  work,  and  instrument  damage  or  serious 
personal injury may be caused.
●  Never attempt to make measurement if any abnormal conditions, 
such  as broken  case  and  exposed  metal parts  are  found  on the 
●  Verify proper operation on  a known source before starting  to use 
the instrument or taking action  as a result of the indication of the 
●  Do  not  install  substitute  parts  or  make  any  modifications  to  the 
instrument.  Return  the  instr
ument  to  your  local  KYORITSU 
distributor  for  repair  or  re-calibration  in  case  of  suspected  faulty 
●  Do not try to replace the batteries if the surface of the instrument 
●  Ensure  that  the  Clamp  sensor  is  disconnected  from  the  object 
under test,  and  that the  instrument is  powered off  when opening 
the Battery compartment cover for battery replacement.
●  This  instrument  is  designed  for  residential,  commercial  or  light 
industry  applications. Accurate  results  may  not  be  obtained  if 
equipments  generating  strong  electromagnetic  interferences 
or  strong  magnetic  fields  due  to  large  currents  exist  in  the 
●  Set the Function switch to the appropriate position before starting 
●  This instrument isn t water proofed. Keep 
●  Be  sure  to  power  off  the  instrument  after  use.  Remove  the 
batteries if the instrument is to be stored and will not be in use for 
●  D o  n ot  e x po s e  t he  in st r um e nt  t o  d ir e ct  s u nl i gh t ,  h ig h 
temperatures, humidity or dew.
●  Use a damp cloth with water or neutral detergent for cleaning the 
instrument. Do not use abrasives or solvents.
● Marks listed below are used on this
User  must  refer  to  the  explanations  in  the  instruction 
Instrument with double or reinforced insulation.
Must wear  insulated gears  such as  a pair  of rubber  gloves 
when  connecting  /  disconnecting  the  sensor  to  /  from  live 
Cros sed-ou t  whee l  bin  symbol  (acco rding  to  WE EE 
Directive: 2002/96/EC) indicating that this electrical product 
may not be treated as household waste, but that it must be 
collected and treated separately.
● Measurement categories(Over-voltage categories)
To  ensure  safe  operation  of  measuring  instruments,  IEC  61010 
establishes  safety  standards  for  various  electrical  environments, 
AT IV, and called measurement categories.
Higher-numbered  categories  correspond  to  electrical  environments 
with  greater  momentary energy,  so  a measuring  instrument  designed 
for  CAT  III  environments  can endure  greater  momentary  energy  than 
O  :  Circuits which  are not  directly connected to  the mains  power 
CAT II  :  Electrical circuits  of equipment connected  to an AC ele
CAT III  :  Primary electrical circuits of the equipment connected directly 
to  the  distribution  panel,  and  feeders  from  the  distribution 
CAT IV :  The circuit from the service drop to the service entrance, and 
to the power meter and primary overcurrent protection device 
● Flexible and light weight Sensor with air core coil
● Wide measuring ranges up to 3000A (30A/ 300A/ 3000A)
● Auto-power-off function
● Designed to meet the following safety standard:
  IEC 61010-1 (CAT III 1000V / CAT IV 600V Pollution degree 2)
● Measuring range and accuracy(23ºC±5ºC, RH 80% or less)
30A 0.00 - 31.49A 1.50 - 30.00A
(At the center of the circle 
formed by the flexible sensor.)
300A 0.0 - 314.9A 15.0 - 300.0A
3000A 0 - 3149A 150 - 3000A
 Crest factor (CF): Full scale CF < 1.6, half scale CF < 3.2.
  Effective input crest values are  2 times of the max values of each √
● Influence of Conductor position
t accuracy is guaranteed when the measured object is 
placed at the center of the clamp sensor.
   The  following  errors  should  be  considered  and  added  to  the 
accuracy depending on the distance from the center position.
Radius 50 mm (ø100) ±2.0%
Radius 75 mm (ø150) ±3.0%
Applicable standards  IEC61010-1, IEC61010-2-030
  CAT III 1000V / CAT IV 600V Pollution degree 2
  IEC61010-2-032 , IEC61326-1(EMC)
Display  Liquid crystal display
Refresh rate  Approx. 2 times per second
Location for use  In-door use, altitude 2000m or les
& humidity  0 to +50  RH80  or less (no condensation)℃ %
& humidity  -10 to +60  RH70  or less (no condensation)℃ %
Power source  Size AAA battery x 2 pcs
  (The use of alkaline LR03 is recommended.)
Approx. 120 hours continuous (with Backlight off)
Low battery warning     appears when the battery voltage drops 
Auto-power-off   Power off function operates in 15 min. after the 
Overload Protection  AC 5000A for 10 sec.
  Add 0.1 x specified accuracy/ ºC
  (above 28ºC or below 18ºC)
Withstand voltage  AC8200V for 5 sec
  (between clamp sensor and enclosure)
Insulation resistance  100M  or more/ 1000VΩ
Conductor size  Max.  150mmФ
Dimension  120(L)x70(W)x26(H)mm
Weight  Approx. 300g (including batteries)
Cable length   Approx.  1.8m(between clamp  sensor  and  main 
Accessory  Carrying case MODEL9174 x 1 pce
  Instruction manual x 1 pce
(1) Checking battery voltage
   Set the Function switch to any position other than the OFF position. 
    When  the  indications  on  the  display  are  clearly  legible  and  the 
" mark is not on, the battery voltage is OK.
    If  the  display  is  blank  or  "
"  mark  is  on,  replace  the  batteries 
according to Section 8: Battery Replacement.
●  When  the  instrument  is  left  powered  on,  the  Auto-power-off 
function  automatically  shuts  the  power  off;  the  display  will  be 
blank even if the Function switch is set to any positions other than 
the OFF position in this state.
   To  power  on  the  instrument,  rotate  the  Function  switch  or  press 
any  of  the  buttons.  If  the  display  is  still  blank,  the  batteries  are 
(2) Checking Function switch position
    Set  the Function  switch  to  the appropriate  range  according  to the 
measurement  purpose.  Confirm  that  the  Data  hold  function  is  not 
6. Operating instructions
●  Never make measurement on a circuit in the following categories;
●  Never  open  the  Bat tery  compartment  c over  while  making 
●  To avoid getting electrical shock by touching the equipment under 
test  or  its  surroundings,  be  sure  to  wear  insulated  protective 
the illustration to the right.
(2)  Clamp onto one conductor under the test, and re-connect the Joint 
connector. Hold the conductor at the center of the Clamp sensor.
(3)  Confirm that the Joint connector on the Clamp sensor is firmly 
●  Jointed part of the Clamp sensor may be disconnected if excessive 
●  Clamp  onto  one  conductor  only;  measurements  cannot  be  made 
when clamping single-phase (2-wi
re) or three-phase (3-wire) at  the 
7-1 Auto-power-off function
      This function is to prevent the battery from being exhausted by the 
instrument being unintentionally left on. 
      The  instrument automatically shifts  to the power-off state  about 15 
min after the last Function switch or other switch operation.
      To  exit from  the Auto-power-off  status, press  any button  or set  the 
Function  switch  to  the  OFF  pos
ition  once,  and  then  set  it  to  the 
  mark  is  displayed  on  the  LCD  when  the Auto-power-off 
[To cancel the Auto-power-off function]
      To  cancel  the  Auto-power-off  function,  hold  down  the  Data  hold 
button and turn the Function switch from OFF position to any other 
      While this  function is  disabled, 
  mark is  not displayed on  the 
[To enable the Auto-power-off function again]
     Turn the Function switch to the OFF, and then set it to any position.
      This is a function to hold measured values on the display.
      Press the Data hold button once to hold the current reading. In this 
data hold state, the reading is held even if input varies. The " " HOLD
  the  LCD.  To  exit  the  data  hold  state,  press  the 
      Press the  Backlight button and  turn on/ off the  LCD backlight. The 
backlight is automatically turned off in 30 sec.
      Displayed  values  can  be  toggled  in  the  following  sequence  by 
pressing the  MIN MAX  button.
      Maximu m  value  ( MAX   appe ars)  -  Mini mum  value  ( MIN  
meas ured  val ue  ( MAX  MI N   bli nks)  - 
Maximum value ( MAX  appears) -…..
      To  disable this  function, hold  down the  MIN MAX  button at  least 2 
sec or rotate the Function switch.
●  Do not try to replace the batteries if the surface of the instrument 
●  Ensure  that  the  Clamp  sensor  is  disconnected  from  the  object 
under test,  and  that the  instrument is  powered off  when opening 
the Battery compartment cover for battery replacement.
●  Never  open  the  Bat tery  compartment  c over  while  making 
● Do not mix new and old batteries or mix different types of batteries.
● Install batteries in correct polarity as marked inside.
Replace  batteries  with  the  new  ones  when  the  empty  battery  mark 
  is  displayed  on  the  LCD.  The  LCD  does  not  show  anything, 
even  the  empty  battery  mark,  when  the  batteries  are  completely 
[ How to replace batteries ]
(1) Power off the instrument.
(2)  Loosen  the  screw  at  the  backside 
of  the  instrument  and  remove  the 
Battery compartment cover.
(3)  Remove  al l  the  old  ba tteries  and 
i ns ta l l  n ew  o n es ,  t wo  si ze  AA A 
batterie s,  in  correct  p olarity.  The 
kal ine  b atte ry  ( LR03)  is 
(4)  Reattach  the  Battery  compartment 
cover and tighten the screw.
Ba y c rt t cover tter ompa men
Kyoritsu  reserves  the  rights  to  change  specifications  or  designs 
described in this manual without notice and without obligations.