Kyoritsu KEW 8178 Manual

Kyoritsu Måling KEW 8178

Læs gratis den danske manual til Kyoritsu KEW 8178 (1 sider) i kategorien Måling. Denne vejledning er vurderet som hjælpsom af 8 personer og har en gennemsnitlig bedømmelse på 4.7 stjerner ud af 4.5 anmeldelser. Har du et spørgsmål om Kyoritsu KEW 8178, eller vil du spørge andre brugere om produktet?

Side 1/1
Instructionmanual
IorLeakageCurrentClampSensor
CLAMPSENSORSeries
KEW8177/8178
1. Safety warnings
KEW 8177/ 8178 Ior leakage clamp sensor (Sensor) has been
designed, manufactured and tested according to IEC 61010: Safety
requirements for Electronic Measuring apparatus, and delivered in
the best condition after passing quality control tests. This instruction
observed by the user to ensure safe operation of the Sensor and to
operating instructions before using the Sensor.
WARNING
Read through and understand instructions contained in this
manual before starting to use the Sensor.
Keep the manual at hand to enable quick reference whenever
necessary.
The Sensor is to be used only in its intended applications.
Understand and follow all the safety instructions contained in the
manual.
It is essential that the above instructions are adhered to. Failure to
follow the above instructions may cause injury and or damage the
Sensor.
The symbol marked on the Sensor, means that the user must refer
to the related parts in the manual for safe operation of the Sensor. It is
essential to read the instructions wherever symbol appears in the
manual.
DANGER : is reserved for conditions and actions that are likely
to cause serious or fatal injury.
WARNING : reserved for conditions and actions that can cause
serious or fatal Injury.
CAUTION : is reserved for conditions and actions that can
cause minor injury or damage the Sensor.
DANGER
Do not attempt to make measurement in the presence of
flammable gasses. Otherwise, the use of the Sensor may cause
sparking, which can lead to an explosion.
Wear insulated protective gears to reduce the risks such as
electrical shock at the workplace.
Never make measurement in CAT III environment if electrical
potentials of 300 V AC or higher exist in the circuit to be tested.
Do not exceed the maximum allowable input of any measuring
range.
The Sensor is to be used only in its intended applications or
conditions.
Otherwise, safety functions equipped with the Sensor will not
work, and Sensor damage or serious personal injury may occur.
Verify proper operation on a known source before starting to use
the Sensor.
Do not make measurement when thunder rumbling. If the Sensor
is in use, stop the measurement immediately and remove the
Sensor from the measured object.
The transformer jaws are designed so as not to short-circuiting
the object under test; however, be especially careful about the
possible shorting where the measured object has uninsulated
conductors.
Never attempt to use the Sensor if it's surface or your hand is wet.
WARNING
Always keep your fingers and hands behind the barrier on the
Sensor to avoid the possible shock hazard.
Never attempt to make any measurement, if any abnormal
conditions are noted, such as broken case, and exposed metal
parts.
Do not install substitute parts or make any modification to the
Sensor.
Return the Sensor to your local distributor for repair or
re-calibration in case of suspected faulty operation.
CAUTION
The Sensor is not dust nor water-proof. Don't use the Sensor at
dusty places or to be spattered.
Take sufficient care to avoid shock when handling the Sensor and
also to prevent foreign substance from being stuck between the
transformer jaws.
Do not step on or pinch the cable to prevent the jacket of cable
from being damaged.
Do not bend or pull the cable of the clamp sensor.
Connect/ disconnect the output terminal without clamping onto a
measured conductor.
Do not expose the Sensor to direct sunlight, high temperature,
humidity or dew.
Never give shocks, such as vibration or drop, which may damage
the Sensor.
Use a damp cloth and detergent for cleaning the Sensor. Do not
use abrasives or solvents.
Safety symbol
User must refer to the explanations in the instruction manual.
Instrument with double or reinforced insulation
Indicates that this instrument can clamp on live bare conductors when
the voltage to be tested is below Circuit Ground-to-Earth voltage
against the indicated Measurement Category.
AC
Crossed-out wheel bin symbol (according to WEEE Directive:
2002/96/EC) indicating that this electrical product may not be
treated as household waste, but that it must be collected and
treated separately.
Measurement Category
To ensure safe operation of measuring instruments, IEC 61010
establishes safety standards for various electrical environments,
categorized as O to CAT IV, and called measurement categories.
Higher-numbered categories correspond to electrical environments
with greater momentary energy, so a measuring instrument designed
for CAT III environments can endure greater momentary energy than
one designed for CAT II.
O
(None, Other)
: Circuits which are not directly connected to the mains
power supply.
CAT II : Electrical circuits of equipment connected to an AC
electrical outlet by a power cord.
CAT III : Primary electrical circuits of the equipment connected
directly to the distribution panel, and feeders from the
distribution panel to outlets.
CAT IV : The circuit from the service drop to the service entrance,
and to the power meter and primary overcurrent
protection device (distribution panel).
2. Features
KEW 8177/ 8178 are clamp sensors to measure Ior leakage
current.
Provides the best phase characteristics when combined and used
with our KEW 5050 Ior leakage logger.
Designed to meet IEC 61010-2-032 (CAT III, Pollution degree 2)
3. Sensor layout
Barrier: provides protection against electrical shock and ensuring the
required minimum air and creepage distances. Always keep
your fingers behind the barrier during a measurement.
4. DIN plug pin assignment
1 : Power
2 : NC (Not connected)
3 : GND pin
4 : SCL pin for memory
5 : Output signal pin
6 : SDA pin for memory
Above figure shows the pin assignment seeing the clamp sensor
from output connector part. The figure of the pin assignment of
connection terminal is symmetrical to above figure.
Power to the Sensor is supplied from KEW 5050 via cable. Connect
a power supply of +5 V across GND pin (no.3) and Power pin (no. 1).
The internal memory stores the information about sensor type,
serial no. and phase correction value. The Sensor is automatically
detected by KEW 5050* by its communication function.
* KEW 5050 is an Ior leakage logger which has a reference voltage
input terminal and four current input terminals and can measure
and record resistive leakage current (Ior) up to four systems.
For its specs and functions, please see the instruction manual for
KEW 5050.
5. Operating instructions
DANGER
The rated ground-to-voltage is CAT III 300 V. Never make
measurement if the potential of the circuit under test exceeds 300 V.
The transformer jaws are designed so as not to short-circuiting
the object under test; however, be especially careful about the
possible shorting where the measured object has uninsulated
conductors.
CAUTION
Take sufficient care to avoid shock, vibration or excessive force
when handling the Sensor. Otherwise, precisely adjusted
transformer jaws will be damaged.
When transformer jaws do not fully close, never try to close them
by force, but make them free to move and try again. If a foreign
substance is stuck in the jaw tips, remove it.
Hold the inserting part (except for the cable) and disconnect the
output terminal from the measuring instrument so as not to cause
a break in the cable.
Do not forcibly open the jaws if they are frozen.
Notes:
The Sensor is dedicated to our KEW 5050 Ior Leakage
Logger. When connecting to our other models, such as KEW
5010/ 5020, the sensor auto-detecting function doesn't work.
Ensure that the transformer jaws are fully closed while clamping
onto a conductor to be tested; otherwise, accurate measurements
cannot be taken. See 6. Specification in this manual for maximum
conductor size.
Sensitive transformer jaws are used for leakage clamp sensor.
Because of the characteristics of split-core type transformer, it is
impossible to eliminate the interference of external magnetic field
completely. If there is a presence of strong magnetic field, use the
Sensor at a distance as far as possible from it. The followings are
the typical things which generate magnetic field.
  
Conductor fed large current
  
Motor
  
Equipment which has magnet
  
Integrating wattmeter
(1) Attach the color markers, supplied with KEW 5050, to clamp
sensors for easy recognition. Colors of the marker are harmonized
with that of current input terminals (red: A1, yellow: A2, blue: A3,
green: A4).
(2) Connect the output terminal to the current input terminal on KEW
5050.
(3) Power on KEW 5050 and wait for a while until the connected
sensor is identified. * Sensor detection can be performed on
Basic setting screen. See 6.2 Basic settings in the manual
for KEW 5050 for details.
(4) Connect the clamp sensor and the voltage test lead properly.
(5) Ensure that the arrow mark on the clamp sensor points towards
load side (towards neutral at earth line measurement). Ensure that
the tips of transformer jaws are firmly closed.
6. Specifications
Model name KEW 8177 KEW 8178
Rated current 10 A (rms) AC (14.1 A peak)
Output voltage 500 mV AC/ 10 A AC (50 mV/ A)
Measuring range 0 10 A AC
Accuracy
(Input: sine wave)
± ±1.0%rdg 0.025 mV (40 - 70 Hz)
± ±4.0%rdg 0.025 mV (30 - 5 kHz, with inputs
of 100mA or more)
Phase characteristics
Within 1.0°±
(45 70 Hz while combining with KEW 5050,
under the input of 10% or more of KEW 5050
leakage current range)
Current consumption 8.6 mA max.
Temperature &
humidity range
(Guaranteed accuracy)
23 5°C, relative humidity 85% or less ±
(no condensation)
Operating temperature
& humidity range
-10 to 50°C, relative humidity 85% or less
(no condensation)
Storage temperature
& humidity range
-20 to 60°C, relative humidity 85% or less
(no condensation)
Max. allowable input*1 100 A (rms) AC, continuous, (40 - 70 Hz)
Output impedance Approx. 100 or less Approx. 60 or lessΩ Ω
Location for use Altitude up to 2000 m, in-door use
Applicable standard
IEC61010-1, IEC61010-2-032
CAT III 300 V Pollution degree 2
IEC 61326-1 (EMC)
Environmental standard EU RoHS directive compliant
Withstand voltage
3470 V AC (rms, 50/ 60 Hz) / 5 sec.
*
Any combination of: engaged Jaws, enclosure,
output terminal
Insulation resistance
50 M or more (@1000 V)Ω
*
Any combination of: engaged Jaws, enclosure,
output terminal
Conductor size Approx. Ø40 mm max. Approx. Ø68 mm max.
Outer dimension
128(L) 81(W) 36(D)mm 186(L) 129(W) 53(D)mm× × × ×
Cable length Approx. 3 m
Weight Approx. 280 g Approx. 560 g
Accessories
Instruction manual,
Carrying case
(KEW 8177: M-9095, KEW 8178: M-9094*2)
*1: Allowable limit values in the case of in correct operations. Accuracy
is not guaranteed for current inputs higher than the rated current.
*2: Carrying case isn't packed with the clamp sensor supplied as
standard accessory.
MEMO
L
N
Source
Load
Connectors on KEW 5050
L1
N
L2
Connectors on KEW 5050
Source
Load
L1/R
L2/S
L3/T
Connectors on KEW 5050
Source
Load
L1/R
L2/S
L3/T
N
Connectors on KEW 5050
Load
Source
92
-
2306
A
11
-
22
DISTRIBUTOR
Incoming wire
:
Device which is
not directly
connected to the
mains power supply
Socket
Interior wiring


Produkt Specifikationer

Mærke: Kyoritsu
Kategori: Måling
Model: KEW 8178

Har du brug for hjælp?

Hvis du har brug for hjælp til Kyoritsu KEW 8178 stil et spørgsmål nedenfor, og andre brugere vil svare dig