[7] AFTER-SALE SERVICE
7-1 Warranty and Provision
Sanwa offers comprehensive warranty services
to its end-users and to its product resellers.
Under Sanwa's general warranty policy, each
instrument is warranted to be free from defects
in workmanship or material under normal use for
the period of one (1) year from the date of
purchase.
This warranty policy is valid within the country of
purchase only, and applied only to the product
purchased from Sanwa authorized agent or
distributor.
Sanwa reserves the right to inspect all warranty
claims to determine the extent to which the
warranty policy shall apply. This warranty shall
not apply to disposables batteries, or any
product or parts, which have been subject to
one of the following causes:
1. A failure due to improper handling or use that
deviates from the instruction manual.
2.
A failure due to inadequate repair or modification
by people other than Sanwa service personnel.
3. A failure due to causes not attributable to this
product such as fire, flood and other natural
disaster.
4. Non-operation due to a discharged battery.
5. A failure or damage due to transportation,
relocation or dropping after the purchase.
7-2 Repair
Customers are asked to provide the following
information when requesting services:
1. Customer name, address, and con tact
information
2. Description of problem
3. Description of product configuration
4. Model Number
5. Product Serial Number
6. Proof of Date-of-Purchase
5-3 Current Measurement (ACA)
Remove the test leads from the measuring
terminals to avoid electric shock.
AC
A
ACA t tcurren measuremen
Remarks:
• Clamp the conductor (cable) to measure at
the center of the trans core (CT).
• Clamp only one cable. If several cables are
clamped together or 2-core or 3-core cords
or cables are clamped, current cannot be
measured accurately.
• The meter may malfunction in places where a
strong magnetic field is present.
• Do not apply voltage and current at the same
time.
5-4 Voltage Measurement (ACV DCV)
Nev er u s e t h e m et er i n p ower li n es
exceeding 600 V to ground.
AC
VDC
V
DC t evol ag
measurement
AC t evol ag
measurement
5-5 Resistance Measurement, Continuity
Check
Never apply a voltage to the input terminals.
Resis anc measurement e t
Continuit checy k
below 30 ohm
Resistor
Continuity
Remarks:
• Reading is 30 ohm or below, buzzer sounds.
• Open voltage: approx. 0.9 V
[6] MAINTENANCE
1. T he following inst ruct ions are very
important for safety. Read this manual
thoroughly to ensure correct maintenance.
2. Calibrate and inspect the meter at least
once a year to ensure safety and maintain
its accuracy.
6-1 Maintenance and Inspection
1) Appearance:
Is the meter not damaged due to falling or
other cause?
2) Test leads:
Are the core wires not exposed at any place
of the test leads?
Is the plug when inserted t o t he input
terminal not loose?
If any of the above problems exists, stop using
the meter and request for repair.
6-2 Calibration and Inspection
For more information, please contact Sanwa’ s
authorized agent / distributor service provider,
listed in our website. See section 7-3.
6-3 Storage
1. The panel and case are not resistant to
volatile solvent and must not be cleaned
with thinner or alcohol.
2. The panel and case are not resistant to
heat. Do not place the meter near heat-
generating devices.
3. Do not store the meter in a place where
it may be subjected to vibration or from
where it may fall.
4. Do not store the meter in places under
direct sunlight, or hot, cold or humid
places or places where condensation is
anticipated.
5. If the meter will not be used for a long
time, remove the batteries.
6-4 Battery Replacement
Batteries when the meter is shipped:
A battery for monitoring has been installed
prior to shipment from the factory. It may be
di sch arged before t h e expirat ion of t h e
described battery life.
*The battery for monitoring is a battery used to
check the functions and performance of the
product.
To avoid electric shock, do not remove the
battery door with an input being applied to
the measuring terminals. Also, before
starting replacement, make sure the power
of the meter is OFF.
− 7 − − 8 − − 9 − − 10 − − 11 −
1. Remove the fixing screw of the battery lid
with a minus screwdriver.
2. Replace both two batteries in the battery
holder with new ones. (Pay attention to their
polarity.)
3. Set and secure the battery door with the
fixing screw as before.
7. Where you purchased the product
Please con tact San wa authorized agen t /
distributor / service provider, listed in our
website, in your country with above information.
An instrument sent to Sanwa / agent / distributor
without above information will be returned to the
customer.
Remarks:
1) Prior to requesting repair, please check the
following: Capacity of the built-in battery, polarity
of installation and discontinuity of the test leads.
2)
Repair during the warranty period: The failed meter
will be repaired in accordance with the conditions
stipulated in 7-1 Warranty and Provision.
3)
Repair after the warranty period has expired: In
some cases, repair and transportation cost may
become higher than the price of the product.
Please contact Sanwa authorized agent /
distributor / service provider in advance.
The minimum retention period of service functional
parts is 6 years after the discontinuation of
manufacture. This retention period is the repair
warranty period. Please note, however, if such
functional parts become unavailable for reasons of
discontinuation of manufacture, etc., the retention
period may become shorter accordingly.
4)
Precautions when sending the product to be
repaired.
To ensure the safety of the product during
transportation, place the product in a box that is
larger than the product 5 times or more in
volume and fill cushion materials fully and then
clearly mark “Repair Product Enclosed” on the
box surface. The cost of sending and returning
the product shall be borne by the customer.
7-3 SANWA web site
https://www.sanwa-meter.co.jp
E-mail: exp_sales@sanwa-meter.co.jp
[8] SPECIFICATIONS
8-1 General Specifications
Operation
method
Double integration
AC measuring Average sensing
Display 6000 counts for A / V
9999 counts for
Resistance
and Continuity check
Sampling rate Approx. 2 times / sec.
Over-range “OL” indication
Range selection Auto
Polarity switching Auto (- indication)
Low Battery
indication
lights when built-in battery
voltage is 2.2 V or below
Max. clamp
conductor
diameter
Φ40 mm
35 mm (open CT size)
Environmental
condition
Altitude up to 2000
meters, indoor use
pollution degree 2
Operating
temperature
5 ºC to 40 ºC.
Humidity range is as
follows (without
condensation):
Max. 80 %RH for
temperatures up to
31 °C decreasing linearly
to 50 % relative humidity
at 40 °C
Storage
temperature
-10 ºC to 40 ºC: ≤80 %RH
(without condensation).
40 ºC to 50 ºC: ≤70 %RH
(without condensation).
(Batteries should be
removed when the
instrument is not to be
used for a long period.)
Power supply LR03 x 2 pcs
Auto Power Off Power Off in about 30
mins. after the last
operation.
Power
consumption
Approx. 70 mW TYP.
Battery life Approx. 42 hours
(backlight off)
Dimensions &
mass
H W D206 x 83 x 38 mm
Approx. 320 g (batteries
included)
Safety standards IEC61010-1 CAT.III 600V,
IEC61010-2-032,
IEC61010-031
EMC Directive IEC61326 (EMC)
RoHS Directive EN50581 (RoHS)
Standard
Accessories
Manual, Test leads
(TL-21a), Carrying case
(C-DCM600)
8-2 Measuring Range and Accuracy
Temperature: 23 °C ± 5 °C, humidity: 80 %RH
max. (no condensation)
rdg (reading): Read value.
dgt (digit): Number of counts of last digit.
ACmA
Function Range Accuracy
ACmA
60.00 mA 50 Hz / 60 Hz
± (1. 2 %rdg + 5 dgt)
40 Hz ~ 400 Hz
± (2.5 %rdg + 10 dgt)
600.0 mA
ACA
Function Range Accuracy
ACA
60.00 A
0 A ~ 200 A
50 Hz / 60 Hz
±
(
1. 2 %rdg + 5 dgt
)
40 Hz ~ 400 Hz
± (2.5 %rdg + 10 dgt)
200.1 A ~ 400 A
50 Hz / 60 Hz
±
(
2.0 %rdg + 5 dgt
)
40 Hz ~ 400 Hz
± (5.0 %rdg + 10 dgt)
400.0 A
Remarks:
• Accuracy is specified at the center of CT.
• Add 0.3 % rdg to the accuracy of 50 Hz in
case of the 50 Hz BPF mode.
• Add 0.3 % rdg to the accuracy of 60 Hz in
case of the 60 Hz BPF mode.
ACV
Function Range Accuracy
ACV 600.0 V ± (1.2 %rdg + 5 dgt)
Remarks:
• Input impedance: Approx 1M ohm
• Accuracy assured frequencies: 40 Hz ~ 400
Hz (sine wave)
DCV
Function Range Accuracy
DCV 600.0 V ± (1.0 %rdg + 2 dgt)
Remarks:
• Input impedance: Approx 1M ohm
Resistance and Continuity check
Function Range Accuracy
999.9 Ω ± (1.0 %rdg + 8 dgt)
Remarks:
• Open voltage: Approx 0.9 V
• Buzzer sounds : Less than 30 ohms
− 6 −
WARNING WARNING
WARNING
WARNING
WARNING
WARNING