Thank you for purchasing the new AC clamp meter. In order to use
this product safely and correctly, please read this manual thoroughly,
especially the Safety Instructions part.
After reading this manual, it is recommended to keep the manual at
an easily accessible place, preferably close to the device, for future
reference.
Limited Warranty and Liability
Uni-Trend guarantees that the product is free from any defect in material
and workmanship within one year from the purchase date. This warranty
does not apply to damages caused by accident, negligence, misuse,
modification, contamination or improper handling. The dealer shall
not be entitled to give any other warranty on behalf of Uni-T rend. If
you need warranty service within the warranty period, please contact
your seller directly.
Uni-Trend will not be responsible for any special, indirect, incidental
or subsequent damage or loss caused by using this device. As some
countries or regions do not allow limitations on implied warranties and
incidental or subsequent damages, the above limitation of liability may
not apply to you.
UT201+/UT202+/UT202A+
AC Clamp Meter User Manual
I. Overview
The UT201+, UT202+, and UT202A+ are true RMS AC clamp meters.
They are designed accord ing to EN61010-1 and CAT II 600V/
CAT III 300V safety standards. These meters come with full-featured
protection which ensures users a safe and reliable measurement
experience. Aside from all the normal features of AC clamp meter,
these meters also include high voltage frequency measurement, fast
capacitance measurement, audio visual NCV detection, and plenty
of additional safety features. UT202A+ also comes with live/neutral
wire detection.
II. Features
• True RMS measurement
Audio visual NCV detection
• Maximum measurable voltage: 600V;
• High voltage frequency range: 10Hz~10kHz
• Current (UT201+/UT202+: 400A, UT202A+: 600A) frequency
response: 50Hz~100Hz; current frequency measurement function
• Large capacitance (4mF) and temperature measurement (UT202+ only)
• Very large capacitance (60mF), low voltage frequency (10MHz),
and live/neutral wire measurement functions (UT202A+ only)
• Large LCD and fast refresh rate (3 times/s)
• Response time for capacitance measurement: less than 3s for 1mF;
about 6s for 10mF; about 8s for 60mF
• Full-featured false detection protection for up to 600V (30kVA) energy
surge; overvoltage and overcurrent alarm functions
• The power consumption of the meter is about 1.8 mA. The circuit
has an automatic power saving function. The consumption in sleep
state is <11uA, which effectively extends the battery life to 400 hours.
Warning: Before using the meter, please read the Safety
carefully. Instructions
III. Accessories
Open the package box and take out the meter. Please double check
whether the following items are missing or damaged.
a) User manual ---------------------------------------------- 1 pc
b) Test leads 1 pair -------------------------------------------------
c) K-type temperature probe (UT202+ only) 1 pc ---------
d) Cloth bag 1 pc --------------------------------------------------
If any of the above is missing or damaged, please contact your supplier
immediately.
IV. Safety Instructions
The meter is designed according to EN61010-1, 61010-2-032/033
and electromagnetic radiation protection EN61326-1 safety standards,
and conforms to CAT II 600V, CAT III 300V, double insulation and
pollution grade II.
Note: In case the meter is not used in acco rdance with the
operating instructions, the protection provided by the meter may be
weakened or lost.
1.Before use, please check if there is any item which is damaged or
behaving abnormally. If any abnormal item (such as bare test lead,
damaged meter casing, broken LCD, etc.) is found, or if the meter
is considered to be malfunctioning, please do not use the meter.
2.Do not use the meter if the rear cover or the battery cover is not
covered up, or it will pose a shock hazard!
3.When using the meter, keep fingers behind the finger guards of the
test leads, and do not touch exposed wires, connectors, unused
inputs, or circuits being measured to prevent electric shock.
4.The function switch should be placed in the correct position before
measurement. It is forbidden to change the position during
measurement to avoid damage to the meter!
5.Do not apply voltage over 600V between any meter terminal and
earth ground to prevent electric shock or damage to the meter.
6. Be cautious when the measured voltage is higher than 60V (DC)
or 30Vrms (AC) to avoid electric shock!
7.Never input voltage or current which exceeds the specified limit. If
the range of the measured value is unknown, the maximum range
should be selected.
8.Before measuring the resistance, diode and continuity online, switch
off the power supply of the circuit, and fully discharge all capacitors
to avoid inaccurate measurement.
Description
Caution, possibility of electric shock
Alternating current
Direct current
Equipment protected throughout by DOUBLE
INSULATION or REINFORCED INSULATION
Earth (ground) TERMINAL
Warning or Caution
VI. General Specifications
1. Max display: 4099 (UT201+/UT202+), 6099 (UT202A+)
2. Polarity display: Auto
3. Overload display: “OL” or “-OL”
4. Low battery indication: The “ ” symbol is displayed.
5. Low battery shutdown prompt: The “Lo.bt” interface appears on the LCD
and lasts for about 10s, the buzzer beeps three times, and the meter
automatically shuts down.
6. Test position error: If the source under test is not placed at the center of
the clamp jaws when measuring current, ±1.0% additional error in reading
will be produced.
7. Drop protection: 1m
8. The maximum size of jaw opening: 28mm in diameter
9. Battery: AAA battery 1.5V×2
10.Auto power off: If there is no operation of the function switch or any
button for 15 minutes, the meter will automatically power off. This function
can be turned off as needed.
11.Dimensions: 215mm×63.5mm×36mm
12.Weight: About 248g (including batteries)
13.Altitude: 2000m
14.Operating temperature and humidity: C~30°C ( 80%RH), 30°C~40°C
( 75%RH), 40°C~50°C ( 45%RH)≤ ≤
15.Storage temperature and humidity: -20°C~60°C ( 80%RH)
16.Electromagnetic compatibility:
RF=1V/m, overall accuracy = specified accuracy + 5% of range
RF>1V/m, no specified calculation
VII. External Structure (Picture 1)
1. NCV sensing end
2. Clamp jaws
3. Hand guard
4. LED indicator
5. Jaw opening trigger
6. Function switch
7. LCD display
8. Function buttons
9. Positive (+) input jack
10.COM (negative -) input jack
VIII. Button Description
1.SELECT Button
In the composite function position, press this button to switch between the
corresponding measurement functions; in the AC/DC/Hz position (UT202+/
UT202A+), short press this button to switch between the AC and DC
functions, and long press (about 2s) this button to enter/exit the Hz
measurement function.
In the NCV/LIVE position (UT202A+ only), short press this button to switch
between the EFHI and EFLo ranges, and long press (about 2s) this button
to enter/exit the LIVE measurement function.
2.HOLD/BACKLIGHT Button
Short press this button to enter/exit the data hold mode, and long press
(about 2s) this button to turn on/o ff the backlight (the backlight will
automatically turn off after 60s).
3.MAX/MIN Button
Short press this button to enter the maximum/minimum measurement mode
and long press this button to exit (only valid for AC/DC voltage, AC current,
resistance and temperature measurement).
4.REL Button (UT202+/UT202A+)
In the capacitance and voltage positions, press this button to store the
current reading as a reference for future readings. When the LCD display
value is reset to zero, the stored reading will be subtracted from the future
readings. Press this button again to exit the relative value mode.
2.AC/DC Voltage and Voltage Frequency Measurement (Picture 3)
1) Insert the red test lead into the “ ” jack, black into
the “ ” jack.COM
2) Turn the function switch to the AC/DC voltage position, and connect
the test leads with the measured load or power supply in parallel.
3.Resistance Measurement (Picture 4)
1) Insert the red test lead into the “ ” jack, black into
the ” jack.COM
2) Turn the function switch to the “Ω” position, press the SELECT
button to select resistance measurement, and connect the test
leads with both ends of the measured resistance in parallel.
Capacitance
Continuity
Diode
Resistance
Note:
• If the measured resistor is open or the resistance exceeds the
maximum range, the LCD will display “OL”.
• Before measuring the resistance online, switch off the power supply
of the circuit, and fully discharge all capacitors to avoid inaccurate
measurement.
• If the resistance is not less than 0.5Ω when the test leads are
short-circuited, please check the test leads for looseness or other
abnormalities.
• Do not input voltage higher than 30V to avoid personal injury.
4.Continuity Test (Picture 4)
1) Insert the red test lead into the “ ” jack, black into
the “ ” jack.
2) Turn the function switch to the “ ” position, press the SELECT
button to select continuity measurement, and connect the test leads
with both ends of the measured load in parallel.
3) Measured resistance <10Ω: The circuit is in good conduction status;
the buzzer beeps continuously. Measured resistance >31Ω: The
buzzer makes no sound.
Note:
• Before measuring the continuity online, switch off the power supply
of the circuit, and fully discharge all capacitors.
• Do not input voltage higher than 30V to avoid personal injury.
5. Diode Test (Picture 4)
1) Insert the red test lead into the “ ” jack, black into
the “ ” jack. The polarity of the red test lead is “+” and that COM
the black test lead is “-”.of
2) Turn the function switch to the “ ” position, press the SELECT
button to select diode measurement, and connect the test leads
with the positive and negative poles of the measured diode.
3) 0.08V reading <1.2V: The buzzer makes one beep indicating
the normality of the diode. Reading <0.08V: The buzzer beeps
continuously indicating the damage of the diode. For the silicon
PN junction, the normal value is generally about 500~800 mV.
9.When the “ ” symbol appears on the LCD, please replace the batteries
in time to ensure measurement accuracy. If the meter is not in use for a
long time, please remove the batteries.
10.Do not change the internal circuit of the meter to avoid damage to the
meter and user!
11.Do not use or store the meter in high temperature, high humidity, flammable,
explosive or strong magnetic field environments.
12.Clean the meter casing with a soft cloth and mild detergent. Do not use
abrasives or solvents!
V. Electrical Symbols
IX. Operating Instructions
1.AC Current/Current Frequency Measurement (Picture 2)
1) Select the AC current range (4A/6A, 40A/60A or 400A/600A)
2) Press the trigger to open the clamp jaws, and fully enclose one conductor.
3) Only one conductor can be measured at a time, otherwise the
measurement reading will be wrong.
AC current
measurement
Current frequency
measurement
Picture 2
Note:
• The current measurement must be taken within 0°C~40°C. Do not
suddenly release the trigger, as the impact will change the reading
for a short time.
• To ensure measurement accuracy, center the conductor in the jaws.
Otherwise, ±1.0% additional error in reading will be produced.
• When the measured current is 400A (UT201+/UT202+)/ 600A ≥ ≥
(UT202A+), the meter will automatically sound an alarm and the
high voltage alarm prompt “ ” will automatically flash.
• If the LCD displays “OL”, it indicates that the current is over range
and there is a danger of damage to the meter.
Note:
• If the diode is open or its polarity is reversed, the LCD will display
“OL”.
• Before measuring the diode online, switch off the power supply of
the circuit, and fully discharge all capacitors.
• Do not input voltage higher than 30V to avoid personal injury.
Red Black
Red Black
Picture 3
Picture 4
Picture 1
Symbol
Note:
Do not input voltage above 600V. Although it is possible to measure
higher voltage, it may damage the meter.
Be cautious to avoid electric shock when measuring high voltage.
When the measured voltage is 30V (AC) or 60V (DC), the LCD ≥ ≥
will display the high voltage alarm prompt
NCV
NCV
NCV
NCVNCV
UT202A+
±( . %+ )1 0 5
UT201+
UT202+
4 000V.
40 00V.
400 0V.
600V
6 000V.
60 00V.
600 0V.
0 001V.
0 01V.
0 1V.
1V
±( . %+ )1 0 5
±( . %+ )0 8 5 600Vrms
UT202A+
UT201+
UT202+
600Vrms
400 0mV.
4 000V.
40 00V.
400 0V.
600V
600 0mV.
6 000V.
60 00V.
600 0V.
0 1mV.
0 001V.
0 01V.
0 1V.
1V
±( . %+ )0 7 3
±( . %+ )0 5 2
±( . %+ )0 7 3
10Hz
10MHz
0 1. ~
99 9%.
%
0 01Hz. ~
0 01MHz.
0 1%.
±( . %+ )0 1 4
±( . + )3 0 5%
600Vrms
UT202A+
UT201+
UT202+
600Vrms
600Vrms
400.0Ω
4.000kΩ
40.00kΩ
400.0kΩ
4.000MΩ
40.00MΩ
600.0Ω
6.000kΩ
60.00kΩ
600.0kΩ
6.000 MΩ
60.00 MΩ
0.1Ω
0.001kΩ
0.01kΩ
0.1kΩ
0.001MΩ
0.01MΩ
±(1.0%+2)
±(0.8%+2)
±(2.0%+5)
UT202A+
UT201+
UT202+
400.0Ω 600.0Ω 0.1Ω
600Vrms
UT202 UT202A+ +
- ℃~ ℃40 40
40 500℃~ ℃
500 1000℃~ ℃
- ℉~ 40 104
104 932℉~ ℉
932 1832℉~ ℉
± ℃4
±( . %+ )1 5 5
±( . %+ )2 0 5
± ℉6
±( . %+ )2 0 6
±( . %+ )2 5 4
1
1
600Vrms
10 NCV.
NCV
4.000nF
40.00nF
400.0nF
4.000uF
40.00uF
400.0uF
4.000mF
6.000nF
60.00nF
600.0nF
6.000uF
60.00uF
600.0uF
6.000mF
60.00mF
0.001nF
0.01nF
0.1nF
0.001uF
0.01uF
0.1uF
0.001mF
0.01mF
±( . %+ )4 0 10
±( . %+ )4 0 5
± %10
0.01Hz~
0.01kHz
EFLo
EFHI
7.Temperature Measurement (UT202+ only, Picture 5)
1) Insert the positive pole of the temperature probe into the
” jack and negative pole into the “ ” jack.COM
2) Turn the function switch to the ” position and the LCD will °C/°F
display room temperature.
3) Fix the temperature probe on the object to be tested, and read the
temperature value of the tested object directly from the display
after a few seconds.
4) Press the SELECT button to switch between °C and °F.
8. Non-contact AC Electric Field Sensing (NCV, Picture 6a )
The electric field sensing sensitivity is divided into two levels (“EFHI”
and “EFLo”). The meter defaults to “EFHI”. Select different sensitivity
levels for measurement according to the intensity of the measured
electric field. When the electric field is around 220V (AC) 50Hz/60Hz,
select "EFHI"; when the electric field is around 110V (AC) 50Hz/60Hz,
select “EFLo”.
1) Turn the function switch to the NCV position.
2) Bring the NCV sensing end of the clamp jaws close to a charged
electric field (socket, insulated wire, etc.). The LCD will display the
segment "-", the buzzer will beep, and the red LED will flash. As
the intensity of the measured electric field increases, the more the
segments (----) are displayed, and the higher the frequency at which
the buzzer beeps and the red LED flashes.
Note:
• Use the NCV sensing end of the clamp jaws to approach the measured
electric field, otherwise the measurement sensitivity will be affected.
• When the measured electric field voltage is 100V (AC), observe
whether the conductor of the measured electric field is insulated to
avoid personal injury.
9.Live/Neutral Wire Measurement (UT202A+ only, Picture 6b)
1) Turn the function switch to the LIVE position.
2) Insert the red test lead into the “ ” jack, make the
suspended, and use the red test lead to touch the black test lead
wire to distinguish the live or neutral wire.socket or bare
3) When the neutral wire or uncharged object is detected, the "----"
state is displayed.
4) When the AC "live wire" (>60 V) is detected, the LCD will display
"LIVE" accompanied by audio/visual indication.
Note:
• When using the LIVE function, in order to avoid the effect of the
interference electric field of the COM input on the accuracy of
distinguishing the live/neutral wire, please move the black test lead
away from the COM input.
• When using the LIVE function, keep your hand away from the meter
casing.
• When the LIVE function is applied to the measurement of dense
high-voltage electric field, the accuracy of the meter to judge the
"live wire" may be unstable. In this case, it should be judged by the
LCD display and sound frequency together.
10. Others
1) Auto power off: During measurement, if there is no operation of the
function switch or any button for 15 minutes, the meter will automatically
shut down to save power. You can wake it up by pressing any button
or restart it after turning the function switch to the OFF position.
To disable the auto power off function, press and hold the SELECT
button in the off state, and then turn on the meter . To resume the
auto power off function, restart the meter after shutdown.
2) Buzzer: When any button is pressed or the function switch is turned,
if it is valid, the buzzer will make one beep (about 0.25s). When
measuring voltage or current, the buzzer will beep intermittently to
indicate the over range.
3) Low battery detection: The battery voltage will be automatically
detected as long as the meter is on. If it is lower than 2.5V, the LCD
will display the “ ” symbol.
4) Low battery shutdown function: When the battery voltage is lower
than 2.4V, the LCD displays the “ ” symbol, the “Lo.bt” interface
appears and lasts for about 10s, the buzzer makes consecutive
beeps three times, and then the meter automatically shuts down
(no interface is displayed).
X. Technical Specifications
Accuracy: ± (a% of reading + b digits), 1 year calibration period
Ambient temperature and humidity: 23°C±5°C; 80%RH
Temperature coefficient: To ensure measurement accuracy, operating
Note:
• The ambient temperature of the
meter should be in the range of
18-28°C, otherwise it will cause
measurement error.
• The positive and negative poles
of the temperature probe should
be properly connected. Do not
measure non-insulated live
objects to avoid incorrect readings.
• Do not input voltage higher than
30V to avoid personal injury.
4 000A.
40 00A.
400 0A.
UT202A+
6 000A.
60 00A.
600 0A.
±( %+ )4 10
±( %+ )2 10
0 001A.
0 01A.
0 1A.
UT201+
UT202+
UT202A+
UT201+
UT202+
420A 620A
±( . %+ )1 0 5
0.1Hz
1.AC Curren t
Range
Resolution Accuracy
Overload Protection
Current frequency
monitoring:
50Hz~100Hz
• For 4A range, open circuit allows least significant digit <3.
Accuracy guarantee range: 1%~100% of range
• The input current amplitude of the current frequency should be >2A.
Frequency response: 50Hz~100Hz
Range
Resolution Accuracy Overload
Protection
2.AC Voltage
Voltage frequency
monitoring: 10Hz~10kHz
• Input impedance: About 10MΩ
• Frequency response: 45Hz~400Hz, true RMS display
Accuracy guarantee range: 1%~100% of range; the input voltage amplitude
of the voltage frequency should be >5V.
• The AC crest factor of a non-sinusoidal wave can reach 3.0 at 4000 counts
while can only reach 1.8 at 6000 counts. The additional error should be
added for the corresponding crest factor as follows:
a) Add 3% when crest factor is 1~2
b) Add 5% when crest factor is 2~2.5
c) Add 7% when crest factor is 2.5~3
3.DC Voltage
Range
Resolution Accuracy Overload
Protection
• For mV range, short circuit allows least significant digit 5.
Accuracy guarantee range: 1%~100% of range
Input impedance: About 10MΩ
4.Frequency/Duty Ratio (UT202A+ only)
Range Resolution Accuracy Overload Protection
1) Measuring sensitivity:
100kHz: 200mVrms input ≤ ≤
amplitude 30Vrms
>100kHz~1MHz: 600mVrms
input amplitude 30Vrms
>1MHz~10MHz: 1Vrms input
amplitude 30Vrms
2) Duty ratio is only applicable to
the measurement of square wave
10kHz; amplitude: 1Vp-p
Frequency 1kHz Duty ratio:
10.0%~95.0%
Frequency >1kHz Duty ratio:
30.0%~70.0%
5. Resistance
Range
Resolution Accuracy Overload
Protection
6. Continuity
Range
Resolution Accuracy Overload
Protection
Open circuit
voltage: About 2.0V
<10 : Consecutive Ω
beeps
>31 : No beepΩ
temperature should be within 18°C ~28°C and the fluctuation range
should be within ±1°C. When the temperature is <18°C or >28°C,
add temperature coefficient error 0.1 x (specified accuracy)/°C.
600Vrms
UT202A+
UT201+
UT202+
4 000V 6 000V 0 001V. . .
7. Diode
Range
Resolution
Accuracy
Overload
Protection
Open circuit voltage:
About 2.2V (UT201+
/UT202+)/3.9V
(UT202A+)
Measurable PN
junction: Forward
voltage drop 2V.
For the silicon PN
junction, the normal
value is generally
about 0.5~0.8V.
8.Capacita nce
Range
Resolution Accuracy Overload
Protection
• Measured value = displayed value - open circuit value of the test
leads (For capacitance 100nF, it is recommended to use “REL”
measurement mode.)
• For capacitance range, open circuit allows least significant digit 20.
9.Temperature(UT202+ only)
Range Resolution Accuracy Overload
Protection
Range Electric field sensing
sensitivity level Accuracy
To sense the wire above 24±6V
and identify whether the mains
socket is charged
To sense the wire above
74V±12V, to identify whether
the mains socket is charged,
or to judge the live/neutral wire
of the socket according to the
intensity of the sensing
Test results may be affected by different socket designs or wire
insulation thickness.
XI. Maintenance
Warning: Before opening the rear cover of the meter, remove the
test leads to avoid electric shock.
1.General Maintenance
1) When the meter is not in use, place the function switch in the OFF
position to avoid continuous consumption of battery energy.
2) Clean the meter casing with a soft cloth and mild detergent.
Do not use abrasives or solvents!
3) The maintenance and service must be implemented by qualified
professionals or designated departments.
2.Battery Replacement (Picture 7)
1) Turn off the meter and remove the test leads from the input terminals.
2) Unscrew the screw of the battery compartment, remove the battery
cover, and replace the 2 standard AAA batteries according to the
polarity indication.
3) Secure the battery cover and tighten the screw.
Red
Picture 7
Picture 6a Picture 6b
Red
Black
Picture 5
6.Capacitance Measurement (UT202+/UT202A+, Picture 4)
1) Insert the red test lead into the “ ” jack, black into
the “ ” jack.COM
2) Turn the function switch to the “ ” position, and connect the test
leads with the measured capacitance in parallel. For capacitance
100nF, it is recommended to use “REL” measurement mode.
Note:
• If the measured capacitor is short-circuited or the capacitance exceeds
the maximum range, the LCD will display “OL”.
• When measuring capacitance >400μF, it may take some time to
steady the readings.
• Before measuring, fully discharge all capacitors (especially for
capacitors with high voltage) to avoid damage to the meter and user.