RIGOL
Programming Guide
RSA5000 Series Real-t ime
Spectrum Analyzer
Jan. , 26 2018
RIGOL TECHNOLOGIES, INC.
RIGOL
RSA5000 Programming Guide I
Guaranty and Declaration
Copyright
© 2017 RIGOL TECHNOLOGIES, INC. All Rights Reserved.
Trademark Inform
ation
RIGOL is a registered trademark of RIGOL TECHNOLOGIES, INC.
Publication
Number
PGD20102-1110
Software
Version
00.01 0 .0
Software upgrade might change or add product features. Please acquire the latest version of the manual
from RIGOL website or contact RIGOL to upgrade the software.
Notices
RIGOL products are covered by P.R.C. and foreign patents, issued and pending.
RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at
the company’s sole decision.
I nformation in this publication replaces all previously released materials.
Information in this publication is subject to change without notice.
RIGOL shall not be liable for either incidental or consequential losses in connection with the furnishing,
use, or performance of this manual, as well as any information contained.
Any part of this document is forbidden to be copied, photocopied, or rearranged without prior written
approval of RIGOL.
Product
Certification
RIGOL guarantees that this product conforms to the national and industrial standards in China as well as
the ISO9001:2008 standard and the ISO14001:2004 standard. Other international standard conformance
certifications are in progress.
Contact
Us
If you have any problem or requirement when using our products or this manual, please contact RIGOL.
E- mail: service@rigol.com
Website: www.rigol.com
RIGOL Chapter 2 Command System
2-10 RSA5000 Programming Guide
:CALCulate:LLINe<n>:FAIL?
Syntax
:CALCulate:LLINe<n>:FAIL?
Description
Queries the measurement results of the selected limit line and its associated trace.
Parameter
Name
Type
Range
Default
<n>
Disc rete
1|2|3|4|5|6
——
Remarks
If you enable the test for the displayed trace and its corresponding limit line, and each displayed trace has
its corresponding limit line enabled, then you can run the command to query whether the trace test passes.
Return Format
The query returns 0 (pass) 1 (fail)or .
:CALCulate:LLINe<n>:FREQuency:CMODe:RELative
Syntax
:CALCulate:LLINe<n>:FREQuency:CMODe:RELative OFF|ON|0|1
:CALCulate:LLINe<n>:FREQuency:CMODe:RELative?
Description
Enables or disables the coupling functions between the selected limit line data points and the center
frequency.
Queries the of the coupling functions between the selected limit line data points and the center status
frequency.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3|4|5|6
——
——
Bool
OFF|ON|0|1
OFF|0
Remarks
O | FF 0: disables the coupling function. That is, "Fixed" is selected under X to CF, and the frequency of the
current limit line editing point i . s not affected by the center frequency
ON|1: enables the coupling function. That is, "Relative" is selected under X to CF, and the amplitude of the
current limit line editing point is the difference between the current point with the current center frequency.
At this time, if the center frequency changes, the position of the current editing point will move left and
right with it.
Return Format
The query returns 0 or 1.
Example
The following command disables the coupling functions between data points of Limit Line 2 and the center
frequency.
:CALCulate:LLINe2:FREQuency:CMODe:RELative OFF or :CALCulate:LLINe2:FREQuency:CMODe:RELative
0
The following query returns 0.
:CALCulate:LLINe2:FREQuency:CMODe:RELative?
RIGOL Chapter 2 Command System
2-12 RSA5000 Programming Guide
:CALCulate:LLINe<n>:TRACe
Syntax
:CALCulate:LLINe<n>:TRACe 1|2|3|4|5|6
:CALCulate:LLINe<n>:TRACe?
Description
Sets the trace to be tested against the current limit line.
Queries the trace to be tested against the current limit line.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3|4|5|6
——
——
Discrete
1|2|3|4|5|6
Refer to "Remarks"
Remarks
Limit Line 1 and 2: The default is Trace 1.
Limit Line 3 and 4: The default is Trace 2.
Limit Line 5 and 6: The default is Trace 3.
Return Format
The query returns 1, 2, 3, 4, 5, or 6.
Example
The following command sets that Trace 2 is tested against Limit Line 3.
:CALCulate:LLINe3:TRACe 2
The following query returns 2.
:CALCulate:LLINe3:TRACe?
:CALCulate:LLINe<n>:TYPE
Syntax
:CALCulate:LLINe<n>:TYPE UPPer|LOWer
:CALCulate:LLINe<n>:TYPE?
Description
Sets the type of the specified limit line.
Queries the type of the specified limit line.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3|4|5|6
——
——
Keyword
UPPer|LOWer
Refer to "Remarks"
Remarks
Limit Line 1, Limit Line 3, and Limit Line 5 belong to the upper type; Limit Line 2, Limit Line 4, and Limit Line
6 belong to the lower type.
Return Format
The query returns UPP or LOW.
Example
The following command sets Limit Line 2 to the upper type.
Chapter 2 Command System RIGOL
RSA5000 Programming Guide 2- 41
:CONFigure Commands
Command List:
:CONFigure?
:CONFigure:ACPower*
:CONFigure:CNRatio*
:CONFigure:DENSity
:CONFigure:DSPEctrogram
:CONFigure:EBWidth*
:CONFigure:HDISt*
:CONFigure:LPSTep
:CONFigure:MCHPower*
:CONFigure:NORMal
:CONFigure:OBWidth*
:CONFigure:PSGRam
: CONFigure:PSPectrum
:CONFigure:PVT
:CONFigure:SANalyzer
:CONFigure:SPECtrogram
:CONFigure:TOI*
:CONFigure:TPOWer*
Remarks:
The commands with * are only available for series that has been installed with advanced the RSA5000
measurement kit option.
:CONFigure?
Syntax
:CONFigure?
Description
Queries the current measurement function.
Return Format
The query returns SAN, TPOW, ACP, MCHP, OBW, EBW, CNR, HARM, , DENS, SPEC, DSPE, PVT, TOI, NORM
PSP, or PSGR. -The explanations for the return results are as follows: SAN (swept SA), TPOW (time domain
power) , MCHP, ACP (adjacent power) ( -multi channel power), OBW , EBW (occupied bandwidth) (emission
bandwidth), CNR -(carrier/noise ratio), HARM (harmonics distortion), TOI (third order intercept), NORM
(Normal), DENS (Density), SPEC (Spectrogram), DSPE (Density Spectrogram), PVT (power versus time),
PSP (PvT ), PSGR (PvT Spectrogram)Spectrum .
RIGOL Chapter 2 Command System
2-42 RSA5000 Programming Guide
:CONFigure:ACPower
Syntax
:CONFigure:ACPower
Description
Sets the analyzer to be in the ACP measurement state.
Remarks
This command is only available for GPSA mode.
Running this command will discontinue the current measurement and restore the specified measurement of
the instrument to the default factory state.
After changing any measurement setup, running the :READ Commands will initialize the measurement, and
you do not have to set the measurement to the default.
:CONFigure:CNRatio
Syntax
:CONFigure:CNRatio
Description
Sets the analyzer to be in the CNR measurement state.
Remarks
This command is only available for GPSA mode.
Running this command will discontinue the current measurement and restore the specified measurement of
the instrument to the default factory state.
After changing any measurement setup, running the :READ Commands will initialize the measurement, and
you do not have to set the measurement to the default.
:CONFigure:DENSity
Syntax
:CONFigure:DENSity
Description
Sets the analyzer to be in the Density measurement state.
Remarks
This command is only available for RTSA mode.
Running this command will discontinue the current measurement and restore the specified measurement of
the instrument to the default factory state.
RIGOL Chapter 2 Command System
2-46 RSA5000 Programming Guide
:CONFigure:SPECtrogram
Syntax
:CONFigure: SPECtrogram
Description
Sets the analyzer to be in the Spectrogram measurement state.
Remarks
This command is only available for RTSA mode.
Running this command will discontinue the current measurement and restore the specified measurement of
the instrument to the default factory state.
:CONFigure:TOI
Syntax
:CONFigure:TOI
Description
Sets the analyzer to be in the TOI measurement state.
Remarks
This command is only available for GPSA mode.
Running this command will discontinue the current measurement and restore the specified measurement of
the instrument to the default factory state.
After changing any measurement setup, running the :READ Commands will initialize the measurement, and
you do not have to set the measurement to the default.
:CONFigure:TPOWer
Syntax
:CONFigure:TPOWer
Description
Sets the analyzer to be in the time domain power measurement state.-
Remarks
This command is only available for GPSA mode.
Running this command will discontinue the current measurement and restore the specified measurement of
the instrument to the default factory state.
After changing any measurement setup, running the :READ Commands will initialize the measurement, and
you do not have to set the measurement to the default.
RIGOL Chapter 2 Command System
2-54 RSA5000 Programming Guide
:DISPlay:VIEW:DENSity:AADJust
Syntax
:DISPlay:VIEW:DENSity:AADJust
Description
Sets the highest density ue to the highest density value found in the current bitmap; sets the lowest h
density hue t zero density value found in the current bitmap.o the lowest non-
Remarks
This command is only available for Density and Density Spectrogram in RTSA mode.
:DISPlay:VIEW:DENSity:CNONlinear
Syntax
:DISPlay:VIEW:DENSity:CNONlinear <real>
:DISPlay:VIEW:DENSity:CNONlinear?
Description
Sets the curve nonlinearity.
Queries the curve nonlinearity.
Parameter
Name
Type
Range
Default
<real>
Consecutive Real Number
- 100 to 100
75
Remarks
This command is only available for Density and Density Spectrogram in RTSA mode.
Within the range between the highest density hue and lowest density hue, setting the curve nonlinearity
can change the gradient among different density hues, making the displayed results move towards either
the higher or lower end of the gradient. Increasing the curve nonlinearity value will compress the colors
towards the higher end of the color bar, and decreasing the nonlinearity value will compress the colors
towards the lower end of the color bar.
Return Format
The query returns the curve nonlinearity value in scientific notation.
Example
The following command sets the curve nonlinearity value to 50.
:DISPlay:VIEW:DENSity:CNONlinear 50
The following query returns 5.000000e+01.
:DISPlay:VIEW:DENSity:CNONlinear?
RIGOL Chapter 2 Command System
2-62 RSA5000 Programming Guide
Remarks
T his command is only available for Density Spectrogram, PvT Spectrum, or PvT Spectrogram in RTSA mode.
Return Format
T he query returns SPEC or PVT.
Example
T he following command selects the PvT window in the PvT Spectrum view.
:DISPlay:WINDow:SELect PVT
The following query returns PVT.
:DISPlay:WINDow:SELect?
:DISPlay:WINDow:TRACe:Y:DLINe
Syntax
:DISPlay:WINDow:TRACe:Y:DLINe <ampl>
:DISPlay:WINDow:TRACe:Y:DLINe?
Description
Sets the position of the display line. Its default unit is dBm.
Queries the position of the display line.
Parameter
Name
Type
Range
Default
<ampl>
Consecutive Real Number
Current amplitude range
- 25 dBm
Remarks
By default, the display line is disabled. When the display line is enabled for the rst time, its position is -25
dBm.
Return Format
The query returns the position of the display line in scientic notation.
Example
The following command sets the position of the display line to - 10 dBm.
:DISPlay:WINDow:TRACe:Y:DLINe 10-
The following query returns -1.000000e+01.
:DISPlay:WINDow:TRACe:Y:DLINe?
RIGOL Chapter 2 Command System
2-72 RSA5000 Programming Guide
:FETCh:EBWidth?
Syntax
:FETCh:EBWidth?
Description
Queries the result of the emission bandwidth measurement.
Remarks
This command is only valid when the emission bandwidth measurement in GPSA mode is enabled.
This command will select data from the latest measurement results, and then transmit the data to the
output buffer.
Return Format
The query returns the emission bandwidth in integer, and its unit is Hz.
Example
The following query returns 5.000000000e+04.
:FETCh:EBWidth?
:FETCh:HARMonics:AMPLitude:ALL?
Syntax
:FETCh:HARMonics:AMPLitude:ALL?
Description
Queries the amplitudes of the rst 10 harmonics. The first harmonic is the fundamental waveform.
Remarks
This command is only valid when the harmonic distortion measurement in GPSA mode is enabled.
The amplitude unit of the return value is consistent with the current Y-axis unit.
This command will select data from the latest measurement results, and then transmit the data to the
output buffer.
Return Format
The query returns the amplitudes of the first 10 harmonics in scientific notation (separated by commas).
If the number of harmonics under measurement is less than 10, the harmonics that are not measured do
not have return values.
Example
The following query returns
- e - -1.692102000 +01, 6.458423000e+01, 7.509421000e+01,-7.924328000e 7.847027000e+01,- +01,-7.885
457000e+01,-7.882358000e 7.921457+01,- 000e 7.923057000e+01, 7.915358000e+01,- - +01.
:FETCh:HARMonics:AMPLitude:ALL?
Chapter 2 Command System RIGOL
RSA5000 Programming Guide 2- 73
:FETCh:HARMonics:AMPLitude? <n>
Syntax
:FETCh:HARMonics:AMPLitude? <n>
Description
Queries the amplitude of the specified harmonic.
Parameter
Name
Type
Range
Default
<n>
Integer
1 to 10
——
Remarks
This command is only valid when the harmonic distortion measurement in GPSA mode is enabled.
The amplitude unit of the return value is consistent with the current Y-axis unit.
The query returns when the harmonic read does not have data.---
This command will select data from the latest measurement results, and then transmit the data to the
output buffer.
Return Format
The query returns the amplitude of the specified harmonic in scientific notation.
Example
The following query returns -1.692102000e+01.
:FETCh:HARMonics:AMPLitude? 1
:FETCh:HARMonics[:DISTortion]?
Syntax
:FETCh:HARMonics[:DISTortion]?
Description
Queries the percentage of the total harmonic distortion.
Remarks
This command is only valid when the harmonic distortion measurement in GPSA mode is enabled.
This command will select data from the latest measurement results, and then transmit the data to the
output buffer.
Return Format
The query returns the percentage of the total harmonic distortion in scientific notation.
Example
The following query returns 1.510000000 02e- .
:FETCh:HARMonics:DISTortion?
RIGOL Chapter 2 Command System
2-90 RSA5000 Programming Guide
:MMEMory Commands
Command List:
:MMEMory:DELete
:MMEMory:LOAD:FMT
:MMEMory:LOAD:LIMit
:MMEMory:LOAD:STATe
:MMEMory:LOAD:TRACe
:MMEMory:LOAD:TRACe:DATA
:MMEMory:MOVE
:MMEMory:STORe:LIMit
:MMEMory:STORe:MTABle
:MMEMory:STORe:PTABle
:MMEMory:STORe:RESults
:MMEMory:STORe:SCReen
:MMEMory:STORe:STATe
:MMEMory:STORe:TRACe
:MMEMory:STORe:TRACe:DATA
Remarks:
The mode name in the following path contains "gpsa" and rtsa" modes. The return path takes "gpsa" as an
example.
:MM EMory:DELete
Syntax
:MMEMory:DELete <file_name>
Description
Deletes a specied le.
Parameter
Name
Type
Range
Default
<file_name>
ASCII String
——
——
Remarks
<file_name> should contain the path and the filename.
This operation fails if the le does not exist. specied
Example
The following command deletes the "state1.sta" le from the "/gpsa/state" folder.
:MMEMory:DELete /gpsa/state/state1.sta


Produkt Specifikationer

Mærke: Rigol
Kategori: Ikke kategoriseret
Model: RSA5032

Har du brug for hjælp?

Hvis du har brug for hjælp til Rigol RSA5032 stil et spørgsmål nedenfor, og andre brugere vil svare dig




Ikke kategoriseret Rigol Manualer

Rigol

Rigol NFP-3 Manual

6 September 2024
Rigol

Rigol PVA7250 Manual

5 September 2024
Rigol

Rigol PHA1150 Manual

5 September 2024
Rigol

Rigol RM-DG Manual

29 August 2024
Rigol

Rigol RM-1-DG1000Z Manual

29 August 2024
Rigol

Rigol PVP2150 Manual

29 August 2024
Rigol

Rigol PVP3150 Manual

28 August 2024
Rigol

Rigol PHA2150 Manual

28 August 2024
Rigol

Rigol RP7150S Manual

28 August 2024
Rigol

Rigol PVP2350 Manual

28 August 2024

Ikke kategoriseret Manualer

Nyeste Ikke kategoriseret Manualer