Læs gratis den danske manual til Rossmax ThermoCal - Professional Testing Device for Thermometer (2 sider) i kategorien Termometer. Denne vejledning er vurderet som hjælpsom af 19 personer og har en gennemsnitlig bedømmelse på 4.6 stjerner ud af 10 anmeldelser. Har du et spørgsmål om Rossmax ThermoCal - Professional Testing Device for Thermometer, eller vil du spørge andre brugere om produktet?

Side 1/2
Instruction Manual
Model: ThermoCal
Professional Testing Device
for Thermometer
www.rossmax.com
OBM_IB_ThermoCal_ver1411
Manufacturer: Rossmax Swiss GmbH,
Tramstrasse 16, CH-9442 Berneck, Switzerland
2. Heating indicator of the water bath does not light up after the calibration temperature is set
Possible causes Actions to take
Water temperature is higher than the set
temperature Cool down the water in the bath
Defective heating indicator Contact Rossmax for service
Defective temperature controller Contact Rossmax for service
Set temperature is too low Set temperature should be at least 5°C higher
than operating temperature
3. Heating indicator of water bath lights up, whereas the water temperature remains low
Possible causes Actions to take
Defective heater circuit Contact Rossmax for service
Defective temperature controller Contact Rossmax for service
4. Water temperature continues to rise and exceed the set temperature
Possible causes Actions to take
Defective temperature controller Contact Rossmax for service
5. Large uctuation in water temperature
Possible causes Actions to take
Water temperature is not stable yet Wait for temperature to stabilize
Set temperature is too close to ambient
temperature Lower the ambient temperature
Defective temperature controller Contact Rossmax for service
Defective water heater Contact Rossmax for service
6. Power and power indicator are both on, whereas the water circulating pump does not start
Possible causes Actions to take
Foreign object clogs inside the pump Remove the foreign object
Loose screw used to secure the propeller Tighten the screw
Defective motor Contact Rossmax for service
7. Noise generated by the stirrer motor
Possible causes Actions to take
Bearing ball oxidation Replace bearing
Bearing aging Replace bearing
Trouble shooting
1. Power is switched on but the power indicator does not light up
Possible causes Actions to take
No power input Check the power source
Loose plug Insert the plug into the socket securely
Device is powered and running but the power
indicator does not light up Replace the power switch
Power wire broke or loose connection Replace the power wire(s)
Tripped fuse Check the fuse and replace it if needed
WARNING: The symbol on this product means that it’s an
electronic product and following the European directive
2012/19/EU the electronic products have to be disposed
on your local recycling centre for safe treatment.


Produkt Specifikationer

Mærke: Rossmax
Kategori: Termometer
Model: ThermoCal - Professional Testing Device for Thermometer

Har du brug for hjælp?

Hvis du har brug for hjælp til Rossmax ThermoCal - Professional Testing Device for Thermometer stil et spørgsmål nedenfor, og andre brugere vil svare dig




Termometer Rossmax Manualer

Termometer Manualer

Nyeste Termometer Manualer

Techno Line

Techno Line WA 1050 Manual

29 Oktober 2025
Techno Line

Techno Line WA 1010 Manual

29 Oktober 2025
Uni-T

Uni-T UT303D+ Manual

13 Oktober 2025
Uni-T

Uni-T UT300S Manual

13 Oktober 2025
Bartscher

Bartscher D5500 Manual

8 Oktober 2025
CDN

CDN IRL500 Manual

8 Oktober 2025
Bartscher

Bartscher EDD 2500 Manual

8 Oktober 2025
Beurer

Beurer OT30 Manual

8 Oktober 2025